Skip to main content
. 2023 Jun 2;10:24. doi: 10.1186/s40662-023-00343-4

Table 3.

Agreement between the two devices for different parameters analyzed

Parameter Mean difference ± SD 95% CI 95% LoA LoA interval
K1 (D)  − 0.011 ± 0.100  − 0.034, 0.011  − 0.208, 0.185 0.393
K2 (D)  − 0.024 ± 0.142  − 0.056, 0.008  − 0.302, 0.254 0.557
CCT (μm) 3.117 ± 4.026 2.199, 4.034  − 4.774, 11.008 15.783
ACD (mm) 0.012 ± 0.054 0.000, 0.024  − 0.093, 0.118 0.212
LT (mm)  − 0.024 ± 0.088  − 0.045, − 0.004  − 0.198, 0.148 0.347
AL (mm) 0.004 ± 0.020  − 0.001, 0.008  − 0.036, 0.044 0.080

K1 = flat keratometry; K2 = steep keratometry; CCT = central corneal thickness; ACD = anterior chamber depth; LT = lens thickness; AL = axial length; SD = standard deviation; CI = confidence interval; LoA = limits of agreement