Skip to main content
. 2023 Jun 2;25(2):19. doi: 10.1007/s10544-023-00657-z

Table 3.

Characterized performance of the portable OoC platform

Category Parameter Performance
Application Channel types 2 (flow)+1 (vacuum)
Fluid type Liquid, Gas and Vacuum
Liquid flow range 1.5 - 68 μLmin-1a
Gas flow range 1.3 - 20.7 mL min-1b
Vacuum range -157 - -556 mbar
Flow profile Fluctuation-free
Architecture Platform size (L × W × H) 29 cm × 24 cm × 22 cm
Weight 4.8 kg
Cost € 5600
Control Function Process control
Controllability Linear
Liquida: 0.8 μLmin-1
Resolution Gasb: 0.2 mL min-1
Vacuum: 3 mbar
Liquida: 0.58 % RSD*
Flow stability Gasb: 0.56 % RSD*
Vacuum: 0.16 % RSD*
Liquida: 4.5 s
Settling time Gasb: 14 s
Vacuum: 2 s
Battery Power consumption Minimum: 0.5 W
Maximum: 18.5 W
Battery worktime Minimum: 3 h
Maximum: 96 h

aDI water is the tested liquid. The value may change if the liquid changes

bThe fluid is the air at a temperature of 20C