Hemodynamic outcomes

Outcomes Baseline 30 d echo p value (Δ baseline to 30 d) 1 y echo p value (Δ baseline to 1 y)
Mean gradient, mm Hg 39.1 ± 18.2 (95) 22.5 ± 8.8 (94) <0.0001 19.7 ± 7.6 (86) <0.0001
Peak gradient, mm Hg 65.2 ± 28.0 (95) 39.0 ± 14.3 (94) <0.0001 34.5 ± 12.8 (86) <0.0001
Aortic valve area, cm2 1.1 ± 0.5 (92) 1.3 ± 0.3 (92) <0.0001 1.4 ± 0.3 (81) <0.0001
Left ventricular ejection fraction, % 59.4 ± 10.6 (95) 58.1 ± 10.6 (92) 0.12 60.8 ± 9.6 (80) 0.15
Dimensionless velocity index 0.25 ± 0.10 (93) 0.31 ± 0.06 (89) <0.0001 0.32 ± 0.06 (80) <0.0001

All values presented as mean ± SD (n). Echocardiography was not available for all patients with follow-up through 1 year. Of available echocardiography, not all hemodynamic outcomes were evaluable. The 3 patients with explanted valves were censored through 1 year.

p values from paired t-tests.