. 2023 Feb 17;24(6):e13926. doi: 10.1002/acm2.13926

TABLE 3.

End‐to‐end testing and its results for three treatment sites of a Rando phantom

Treatment site Prescribed dose (PD) to treatment depth (cGy) Treatment depth (mm) Energy (kV) Applicator size (cm) Treatment time (min) Expected dose on surface (cGy) Measured dose on surface (cGy) Dose diff. (%) between expected and measured
Chin 500 1.0 50 2.0 0.72 603 606.24 0.54
Cheek 500 1.5 70 2.5 0.84 580 577.91 0.36
Forehead 500 2.0 100 3.0 0.81 557 557.29 0.05