Skip to main content
. 2023 May 30;2023:9624637. doi: 10.1155/2023/9624637

Table 3.

Analytical figures of merit for the determination of Cd and Pb in swimming pool waters by GF AAS after preconcentration on polyurethane foam.

Parameter Pb(II) Cd(II)
Concentration range 1.8–40 μg·L−1 0.06–3 μg·L−1
 LOD 0.5 μg·L−1 0.02 μg·L−1
 LOQ 1.8 μg·L−1 0.06 μg·L−1
Calibration equation A = 0.0069 (Pb (μg L−1)) + 0.0047 A = 0.16 (Cd (μg L−1)) + 0.0018
r2 0.9976 0.9996
 Intermediary precision 7.3% 6.1%