Table 3. Comparison among FGM and SMBG.
| Clarke error grid analysis | Accurate % | Acceptable (benign) % | Erroneous % | Pearson correlation + linear regression FGM x SMBG |
|---|---|---|---|---|
| Overall (n = 26 patients/1,705 measurements) |
96.60 | 0.76 | 2.63 | 0.90 (<0.001) 0.87 {0.85-0.89} |
| Glucose range < 70 mg/dL (N = 25 patients/146 measurements) |
95.38 | 0.51 | 4.10 | 0.16 (=0.02) 0.65 {0.08-1.23} |
| Glucose range 70-180 mg/dL (N = 26 patients/813 measurements) | 97.57 | 0.11 | 2.30 | 0.68 (<0.001) 0.82 {0.76-0.88} |
| Glucose range > 180 mg/dL (N = 26 patients/746 measurements) |
98.12 | 1.73 | 0.15 | 0.74 (<0.001) 0.75 {0.70-0.80} |
FGM: flash glucose monitoring (FreeStyle Libre); SMBG: self-monitoring blood glucose. [97.5% CI], (p value).