. 2022 Dec 1;67(3):289–297. doi: 10.20945/2359-3997000000540

Table 3. Comparison among FGM and SMBG.

Clarke error grid analysis Accurate % Acceptable (benign) % Erroneous % Pearson correlation + linear regression FGM x SMBG
Overall
(n = 26 patients/1,705 measurements)
96.60 0.76 2.63 0.90 (<0.001)
0.87 {0.85-0.89}
Glucose range < 70 mg/dL
(N = 25 patients/146 measurements)
95.38 0.51 4.10 0.16 (=0.02)
0.65 {0.08-1.23}
Glucose range 70-180 mg/dL (N = 26 patients/813 measurements) 97.57 0.11 2.30 0.68 (<0.001)
0.82 {0.76-0.88}
Glucose range > 180 mg/dL
(N = 26 patients/746 measurements)
98.12 1.73 0.15 0.74 (<0.001)
0.75 {0.70-0.80}

FGM: flash glucose monitoring (FreeStyle Libre); SMBG: self-monitoring blood glucose. [97.5% CI], (p value).