Table 2.

The wavelength used with ICP-OES, instrumental detection limit (IDL), limit of detection (LOD), limit of quantification (LOQ), correlation coefficient, and calibration equations

Metal Wavelength IDL (mg/L) LOD (mg/L) LOQ (mg/L) Correlation coefficient Calibration equation
Na 589.592 0.069 2.16 7.22 0.9988 I = 9245C-7006.1
Mg 285.213 0.002 0.55 1.83 0.9999 I = 1575000C-287546.9
Ca 317.933 0.010 2.92 9.75 0.9990 I = 641100C-632445.1
Mn 257.610 0.001 0.54 1.79 0.9990 I = 4813000C-858583
Fe 238.204 0.004 0.67 2.25 0.9996 I = 590900C-131479.4
Cr 267.716 0.007 0.44 1.48 0.9991 I = 640200C-92120
Co 228.616 0.007 0.31 1.03 0.9994 I = 209100C-20260.1
Ni 231.604 0.015 0.39 1.30 0.9994 I = 185200C-20700.7
Cu 327.393 0.009 0.25 0.82 0.9996 I = 822900C-66599.7
Zn 206.200 0.005 0.75 2.50 0.9980 I = 96750C-24088.4
Cd 228.802 0.030 0.15 0.51 0.9995 I = 353000C-14855.8
K 766.490 6.78 21.3- 0.9968 I = 4614C + 3950.7

*I is emission intensity and C is concentration