Table 2.
IGRA performance using different cutoffs.
| Cutoff ≥100 mIU/mL | Cutoff ≥147 mIU/mL | Cutoff ≥200 mIU/mL | |
|---|---|---|---|
| Sensitivity % (95% CI) | 89.4% (81.5–94.3) | 87.2% (79.0–92.5) | 81.9% (72.9–88.4) |
| Specificity (95% CI) | 84.6% (66.5–93.8) | 92.3% (75.9–97.9) | 92.3% (75.9–97.9) |
| Negative predictive value (95% CI) | 68.8% (51.4–82.1) | 66.7% (50.3–79.8) | 58.5% (43.4–72.2) |
| Positive predictive value (95% CI) | 95.5% (88,9–98.2) | 97.6% (91.7–99.3) | 97.4% (91.2–99.3) |
| Accuracy (95% CI) | 88.3% (81.4–92.9) | 88.3% (81.4–92.9) | 84.2% (76.6–89.6) |