Table 2.
The accuracy of ethanol, isopropyl alcohol, and methanol at low, medium, and high-quality control levels for liquid and gel hand sanitisers (established via spike recovery studies).
Form | Analyte | Spike concentration (% v/v) | Mean recovery (%) ± 95% CI | RSD (%) | n |
---|---|---|---|---|---|
Liquid | Ethanol | 0.480 | 101.2 ± 0.004 | 0.011 | 6 |
0.750 | 100.2 ± 0.009 | 0.015 | 6 | ||
0.900 | 99.5 ± 0.017 | 0.023 | 6 | ||
IPA | 0.480 | 100.5 ± 0.003 | 0.008 | 6 | |
0.750 | 100.2 ± 0.005 | 0.008 | 6 | ||
0.900 | 99.5 ± 0.006 | 0.009 | 6 | ||
Methanol | 0.0120 | 98.5 ± 1.292 | 1.640 | 6 | |
0.0375 | 99.9 ± 0.687 | 0.859 | 6 | ||
0.0500 | 101.5 ± 0.741 | 0.912 | 6 | ||
Gel | Ethanol | 0.480 | 97.7 ± 0.004 | 0.011 | 6 |
0.750 | 101.0 ± 0.003 | 0.006 | 6 | ||
0.900 | 99.1 ± 0.008 | 0.011 | 6 | ||
IPA | 0.480 | 97.9 ± 0.002 | 0.005 | 6 | |
0.750 | 99.8 ± 0.003 | 0.005 | 6 | ||
0.900 | 99.3 ± 0.006 | 0.009 | 6 | ||
Methanol | 0.0120 | 97.3 ± 0.624 | 0.801 | 6 | |
0.0375 | 101.8 ± 0.435 | 0.534 | 6 | ||
0.0500 | 100.5 ± 1.077 | 1.340 | 6 |