Table 2.

Index procedure–related characteristics in patients with and without periprocedural cardiac tamponade

Baseline characteristics Tamponade group (n = 200) Control group (n = 400) P value
Type of arrhythmia to be treated, n (%) a 1.000
Atrial fibrillation 133 (66.5) 266 (66.5)
VES/VT 29 (14.5) 58 (14.5)
AVRT 12 (6.0) 24 (6.0)
AVNRT 11 (5.5) 22 (5.5)
Atrial tachycardia excluding CTI-dependent flutter 7 (3.5) 14 (3.5)
His ablation 4 (2.0) 8 (2.0)
CTI-dependent flutter 4 (2.0) 8 (2.0)
Energy delivery, n (%) b 0.065
RF energy 165 (82.5) 343 (85.8)
Cryo energy 12 (6.0) 34 (8.5)
RF and cryo energy 2 (1.0) 3 (0.8)
Unclassified 21 (10.5) 20 (5.0)
Procedure time (min) c 188.8 +/78.5 173.0 +/69.5 0.015
a

Matched variable.

b

Data available in 93% of all patients.

c

Data available in 96% of all patients.

VES, ventricular extrasystole; VT, ventricular tachycardia; AVNRT, atrioventricular node re-entry tachycardia; AVRT, atrioventricular re-entrant tachycardia; CTI, cavotricuspid isthmus; RF, radiofrequency.