Table 2.

Mean values of end-diastolic and end-systolic volumes (EDV and ESV) and corresponding ejection fractions (EF) for the modalities MR, and PET reconstructed with the three methods of cardiac ECG gating: STD: standard ECG gating method, STD-BR: standard - beat rejection ECG gating method, FW: fixed width ECG gating method

Modality EDV (mL) ESV (mL) EF (%)
MR 160 ± 35 82 ± 33 50 ± 14
PET-STD 140 ± 41* 80 ± 33 43 ± 10*
PET-STD-BR 145 ± 40* 78 ± 35 47 ± 11
PET-FW 145 ± 43* 81 ± 37 45 ± 10*

*Stands for statistically significant differences against MR values (P value < .05)