Skip to main content
. 2023 Apr 7;15(5):2485–2496. doi: 10.21037/jtd-22-1546

Table 3. Diagnostic yield, procedure time, and complication rate of US-guided PCNB for subpleural lung lesions.

Variable Total (N=572) Small (N=72) Medium (N=206) Large (N=294) P
Sample adequacy 0.307
   Adequate 550 (96.2) 67 (93.1) 198 (96.1) 285 (96.9)
   Inadequate 22 (3.8) 5 (6.9) 8 (3.9) 9 (3.1)
Success rate 0.079
   Success 474 (82.9) 54 (75.0) 168 (81.6) 252 (85.7)
   Fail 98 (17.1) 18 (25.0) 38 (18.4) 42 (14.3)
Accuracy 0.366
   Accurate 517 (90.4) 62 (86.1) 189 (91.7) 266 (90.5)
   Inaccurate 55 (9.6) 10 (13.9) 17 (8.3) 28 (9.5)
Procedure time, min 10.0±4.7 14.8±6.1 9.8±4.5 8.9±3.7 <0.001*
Complication 27 (4.7) 8 (11.1) 13 (6.3) 6 (2.0) <0.001*
   Pneumothorax 19 (3.3) 6 (8.3) 10 (4.9) 3 (1.0)
   Hemorrhagic 8 (1.4) 2 (2.8) 3 (1.4) 3 (1.7)

Except where otherwise indicated, data are numbered with percentage in parentheses. *, statistically significant. US, ultrasound; PCNB, percutaneous core needle biopsy.