. 2023 Apr 11;11(3):e04044-22. doi: 10.1128/spectrum.04044-22

TABLE 5.

Reproducibility of BioCode RPPa

Pathogen No. of samples detected/tested (% agreement) by site and test
Total no. of samples detected/tested (%) (95% CI)
Site 1, easyMAG Site 2, MagNA Pure Site 3, easyMAG
Adenovirus 60/60 (100) 60/60 (100) 60/60 (100) 180/180 (100) (97.9–100.0)
Coronavirus NL63 60/60 (100) 60/60 (100) 60/60 (100) 180/180 (100) (97.9–100.0)
Human metapneumovirus 60/60 (100) 60/60 (100) 60/60 (100) 180/180 (100) (97.9–100.0)
Human rhinovirus/enterovirus 60/60 (100) 60/60 (100) 60/60 (100) 180/180 (100) (97.9–100.0)
Influenza A H3b 59/60 (98.3) 60/60 (100) 60/60 (100) 179/180 (99.4) (96.9–99.9)
Influenza Bb 59/60 (98.3) 60/60 (100) 60/60 (100) 179/180 (99.4) (96.9–99.9)
Parainfluenza virus 2 60/60 (100) 60/60 (100) 60/60 (100) 180/180 (100) (97.9–100.0)
Parainfluenza virus 3 60/60 (100) 60/60 (100) 60/60 (100) 180/180 (100) (97.9–100.0)
Respiratory syncytial virusb,c 58/60 (96.7) 60/60 (100) 60/60 (100) 178/180 (98.9) (96.0–99.7)
Bordetella pertussis 60/60 (100) 60/60 (100) 60/60 (100) 180/180 (100) (97.9–100.0)
Chlamydia pneumoniae 60/60 (100) 60/60 (100) 60/60 (100) 180/180 (100) (97.9–100.0)
Mycoplasma pneumoniae 60/60 (100) 60/60 (100) 60/60 (100) 180/180 (100) (97.9–100.0)
a

Reproducibility was assessed with a panel of samples at 3 different sites, 2 with easyMAG and 1 with MagNA Pure 96 extraction systems.

b

Low positive, 1.5 times LoD not detected.

c

Medium positive, 3 times LoD not detected.