TableĀ 8.

Reproducibility and intraclass correlation of in-house assaysa

In-House developed SARS-CoV-2 serological assay reproducibility and intraclass correlation coefficient
Assay ID Antibody Within day (%) Across day (%) Overall (%) Intraclass correlation coefficient (ICC)
11 Pan Ig 9.9 12.1 15.6 0.925
12 Pan Ig 8.7 10.8 13.9 0.896
13 Pan Ig 8.3 12.6 15.1 0.936
14 Pan Ig 12.3 15.6 19.9 0.954
15 Pan Ig 8.6 14.6 16.9 0.789
16 Pan Ig 6.6 22.5 23.4 0.964
17 Pan Ig 3.1 9.2 9.7 0.976
18 IgG 12.4 30.2 32.7 0.946
19 IgG 16.3 16.2 23.0 0.980
20 IgG 6.4 25.7 26.5 0.983
21 IgG 8.7 23.1 24.6 0.976
22 IgG 15.4 12.8 20.0 0.986
23 IgG 10.5 10.8 15.0 0.989
24 IgG 8.0 16.4 18.2 0.980
a

Variability above 20% is highlighted in bold. Data analysis was performed prior to harmonization.