Key imaging parameters for the 0.55 T fetal MRI protocol as implemented on a 0.55 T scanner

Sequence TR (ms) TE (ms) FA (deg) BW (Hz/Px) Slice thickness (mm) Slice gap (mm) Avg Number of slices FOV (mm2) Acquisition matrix Voxel size (mm3) Partial Fourier Acquisition time
Per slice (s) Total
HASTE Body 2000 98 180 251 3 0 1 30 350 × 350 224 × 224 1.6 × 1.6 × 3.0 4/8 2.0 1:00
Brain 2000 98 180 251 3 0 1 30 300 × 300 224 × 224 1.3 × 1.3 × 3.0 4/8 2.0 1:00
bSSFP Body 6.02 3.01 120 250 3 0 1 30 350 × 350 192 × 192 1.8 × 1.8 × 3.0 None 1.5 0:45
Brain 6.02 3.01 120 250 3 0 1 30 300 × 300 192 × 192 1.6 × 1.6 × 3.0 None 1.5 0:45
Real-time cine 570 1.82 70 558 10 - 1 1 350 × 350 128 × 128 2.8 × 2.8 × 10 None 0.6 -
T1w GRE 113 2.43 60 250 4 0 1 20 295 × 350 138 × 192 1.8 × 1.8 × 4.0 None - 0:16
DWI (b = 0.700) 2500 74 - 1510 4 0 1 20 294 × 330 82 × 92 3.6 × 3.6 × 4.0 6/8 - 0:30