. 2023 Jun;37(6):641–646. [Article in Chinese] doi: 10.7507/1002-1892.202303123

表 2.

Comparison of radiological results between two groups (n=20)

两组患者影像学结果比较(n=20)

指标
Index
Corail 组
Corail group
ABG Ⅱ组
ABG Ⅱ group
效应值(95%CI
Effect value (95%CI)
P
P value
X线片征象
 轻度应力遮挡(1~2级)(是/否,例) 8/12 12/8 RR=0.44(0.13,1.58) 0.206
 宽度<1 mm透亮线(是/否,例) 1/19 4/16 RR=0.21(0.02,2.01) 0.342
 点焊征(是/否,例) 15/5 14/6 RR=1.29(0.32,5.18) 0.723
 股骨柄末端皮质增厚(是/否,例) 4/16 3/17 RR=1.42(0.27,7.34) 1.000
 基座征(是/否,例) 12/8 3/17 RR=8.50(1.86,38.82) 0.003
 骨质溶解(是/否,例) 1/19 1/19 RR=1.00(0.06,17.18) 1.000
 异位骨化(是/否,例) 0/20 6/14 RR=0.70(0.53,0.93) 0.020
 股骨柄下沉距离(Inline graphic,mm) 1.26±1.22 3.06±3.05 MD=−1.80(−3.24,−0.36) 0.019
 股骨柄下沉速度(Inline graphic,mm/年) 0.14±0.14 0.25±0.26 MD=−0.11(−0.24,0.02) 0.104
假体填充率(Inline graphic,%)
 小转子处 0.72±0.08 0.75±0.07 MD=−0.03(−0.08,0.02) 0.215
 小转子下2 cm 0.80±0.08 0.80±0.09 MD=0.00(−0.05,0.05) 1.000
 小转子下7 cm 0.79±0.12 0.76±0.08 MD=0.03(0.03,0.09) 0.358
 整体 0.29±0.05 0.37±0.08 MD=−0.08(−0.12,−0.04) <0.001
CT观测假体对线误差
 冠状位对线(Inline graphic,°) 1.35±1.04 2.25±1.51 MD=−0.90(−1.70,−0.10) 0.034
 冠状位对线>3°(是/否,例) 3/17 5/15 RR=0.53(0.11,2.60) 0.695
 矢状位对线(Inline graphic,°) 2.50±1.91 2.05±1.70 MD=0.45(−0.67,1.57) 0.427
 矢状位对线>3°(是/否,例) 5/15 3/17 RR=1.89(0.36,9.27) 0.695