Skip to main content
. 2023 Jun 18;24:34. doi: 10.1186/s12868-023-00805-4

Table 1.

Resistance to flow measurements

Type of needle Measurement (M) Total number of drops Total weight of drops (mg) Measurement time (s) Flow rate (cmH2O) R (cm.s/ml) AVG + SD
22G Pajunk Sprotte M1 10 573.60 25.5 0.0225 667.5 667.9 ± 0.91
M2 10 574.20 25.6 0.0224 668.6
M3 10 590.80 26.3 0.0225 667.7
22G Braun Spinocan M1 10 549.60 51.8 0.0106 1415.0 1411.3 ± 3.49
M2 10 539.50 50.7 0.0106 1410.8
M3 10 554.50 52.1 0.0107 1408.0
22G M. Schilling M1 9 480.80 32.6 0.0147 1017.1 1028.7 ± 11.3
M2 9 508.60 34.9 0.0146 1029.3
M3 9 507.80 35.2 0.0144 1039.8