Table 5.
Chromatographic methods for the analysis of the drug substance in different biological matrices.
| Matrix | Objective of analysis | Stationary phase | Mobile phase | Gradient | Detection | Flow rate | Column temperature | Reference | |
|---|---|---|---|---|---|---|---|---|---|
| Urine (Human) | Drug content | Zorbax SB-C8 Column 2.1 × 100mm,1.8 μm  | 
Mobile phase A-10 mM ammonium formate and acetic acid (pH 4) Mobile B- Acetonitrile  | 
Time | %B | Triple quadrupole mass spectrometer (Waters) Positive mode  | 
400 μL/min | 20 °C | [31] | 
| 10 | |||||||||
| 8 | 55 | ||||||||
| 8.1 | 100 | ||||||||
| 8.6 | 10 | ||||||||
| 10.6 | 10 | ||||||||
| Urine (human) | Drug content | Zorbax XDB-C8 Column 2.1 × 150mm,5.0 μm  | 
Mobile phase A-10 mM ammonium formate and acetic acid (pH 4) Mobile B- Acetonitrile  | 
Time | %B | TSQ Quantum Ultra triple quadrupole mass spectrometer With positive mode  | 
250 μL/min | ||
| 0 | 40 | ||||||||
| 5 | 90 | ||||||||
| 9.5 | 90 | ||||||||
| 9.6 | 40 | ||||||||
| 15 | 40 | ||||||||
| Plasma and urine (human) | Drug and metabolite analysis | Biphenyl (100 × 2.1 mm, 2.7 μm | Mobile phase A-25 mM ammonium formate with 0.1% formic acid Mobile B- Acetonitrile | Time | %B | Triple quadrupole mass spectrometer | 400 μL/min | 40 °C | [7] | 
| 0 | 14 | ||||||||
| 0.5 | 70 | ||||||||
| 6.5 | 70 | ||||||||
| 6.6 | 95 | ||||||||
| 7.5 | 95 | ||||||||
| 7.6 | 14 | ||||||||
| 9.0 | 14 | ||||||||
| Human hepatocyte | Metabolites study | Waters, BEH C18 Column 2.1 × 50mm,1.7 μm | Mobile phase A-5 mM ammonium acetate pH 3.5 Mobile B- Acetonitrile | Time | %B | Water Xevo TQ-S triple quadrupole mass spectrometer | 300 μL/min | 50 °C | [33] | 
| 0 | 0.5 | ||||||||
| 0.5 | 5 | ||||||||
| 6.0 | 50 | ||||||||
| 8.0 | 50 | ||||||||
| 10.0 | 70 | ||||||||
| 12.0 | 95 | ||||||||
| 14 | 95 | ||||||||
| 16.5 | 5 | ||||||||
| Liver microsome and fungal sample | Acquity UPLC BEH C18 Column 2.1 × 100 mm,1.7 μm | Mobile phase A-0.1% Formic acid B- methanol | Time | %B | 500 μL/min | 65 °C | |||
| 0 | 5 | ||||||||
| 1.0 | 5 | ||||||||
| 6.0 | 95 | ||||||||
| 8.0 | 95 | ||||||||
| 8.1 | 5 | ||||||||
| 10.0 | 5 | ||||||||
| Plasma and urine(mice) | Metabolites study | Acquity HSS T3; C18, 150mmx2.1 mm,1.8 μm | Mobile phase A-0.01% Formic acid B- Acetonitrile with 0.01% Formic acid | Time | %B | Waters synapt G2-Si QTOF MS with ESI | 400 μL/min | 40 °C | [34] | 
| 0 | 1 | ||||||||
| 1.0 | 1 | ||||||||
| 8.0 | 99 | ||||||||
| 13 | 85 | ||||||||
| 10.0 | 1 | ||||||||
| Plasma (human) | Pharmacokinetics and drug-drug interaction study | C18 Column | Mobile phase A-0.1% Formic acid B- acetonitrile 0.1% Formic acid | Isocratic – A: B (45:55) | [37] | ||||
| Urine | Antidoping analysis | Synchronis C18 (100 × 2.1 mm, 1.7 μm) | Mobile phase A-0.2% Formic acid B- acetonitrile 0.2% Formic acid | Time | %B | Thermo-scientific Q Exactive plus tandem mass spectrometer | 500 μL/min | [38] | |
| 0 | 2 | ||||||||
| 0.5 | 2 | ||||||||
| 8.5 | 95 | ||||||||
| 10.0 | 95 | ||||||||
| 12.0 | 2 | ||||||||
| Urine (human) | Simultaneous analysis of drug | Supelco AscentisR C18 (150 mm × 2.1mm × 2.7 μm)  | 
Mobile phase A-0.1% Formic acid B- acetonitrile 0.1% Formic acid | Time | %B | QExactive focus benchtop Orbitrap-based mass spectrometer | 250 μL/min | 30 °C | [39] | 
| 0 | 5 | ||||||||
| 7 | 65 | ||||||||
| 8 | 100 | ||||||||
| 11.0 | 100 | ||||||||
| 13.0 | 5 | ||||||||
| Urine (Horse) | Metabolites study | Eclipse plus C18 (4.6 × 150mm, 3.5 μm) | Mobile phase A-5Mm ammonium acetate or 0.2% formic in water B- acetonitrile | Time | %B | QExactive high-resolution accurate mass spectrometer | 600 μL/min | [40] | |
| 0 | 2 | ||||||||
| 5 | 95 | ||||||||
| 8.0 | 95 | ||||||||
| 8.5 | 50 | ||||||||
| 13.5 | 50 | ||||||||
| 14.0 | 50 | ||||||||
| 15.5 | 2 | ||||||||
| Liposome formulation | HPLC assay | Supelco Discovery BIO wide pore C5, 250mmx4.6 mm,5 μm | Mobile phase A-5% acetonitrile and 0.1% trifluoroacetic acid in water Mobile phase B-95% acetonitrile and 0.1% trifluoroacetic acid in water  | 
Time | %B | UV 348 nm | 1 mL/min | [41] | |
| 0 | 30 | ||||||||
| 20 | 100 | ||||||||
| 21 | 30 | ||||||||
| Active pharmaceutical substance | Purity analysis | Kinetex C18 150mmX3.0 mm,2.6 μm  | 
Mobile phase A-10 mM ammonium formate pH 6.3 Mobile Phase B-acetonitrile  | 
Time | %B | LTQ XL Orbitrap Mass spectrometer (Thermo) | 0.6 mL/min | [42] | |
| 0 | 30 | ||||||||
| 4 | 30 | ||||||||
| 18 | 100 | ||||||||
| 23 | 100 | ||||||||
| 25 | 30 | ||||||||
| 30 | 30 | ||||||||
| Supelco- Ascentis Express C18 | Mobile phase A-0.01 M pH 3.0 Phase B-Acetonitrile  | 
Time | %B | UV 225 nm | |||||
| 0 | 10 | ||||||||
| 1 | 10 | ||||||||
| 10 | 90 | ||||||||
| 12 | 90 | ||||||||
| 13 | 10 | ||||||||
| 14 | 10 | ||||||||