Skip to main content
. 2023 Jun 3;9(6):e16987. doi: 10.1016/j.heliyon.2023.e16987

Table 2.

Influence of the make-up solvent on the MS detection in SFC-MS.

Entry make-up solvent composition additional ion
m/z (theoretical)
adduct formed
m/z (theoretical)
ion found
1 MeOH:H2O (9:1, v:v)
+0.1% HCOOH
HCOO
45.0
[M + HCOO]-
325.0
324.9
2 MeOH:H2O (9:1, v:v)
+0.1% CH3COOH
CH3COO
59.0
[M + CH3COO]-
339.0
339.0
3 MeOH:H2O (9:1, v:v)
+0.1% NH4OH
CH3OCOO
75.0
[M + CH3OCOO]-355.0 355.0