TABLE 1.
Key features of the automated RPR tests
| Parameters |
Automated RPR |
||
|---|---|---|---|
| BioPlex RPR |
AIX 1000 |
ASI evolution |
|
| Turnaround (tests/hour)a | 200 | 192 | 190 |
| Specimen vialb | |||
| Type | Polystyrene / polypropylene | Polystyrene / polypropylene | Polystyrene / polypropylene |
| Dimension (mm) | 12 × 75 | 16 × 100 | 12 × 75 |
| Specimen volb | |||
| Qualitative / Screening (μL) | 240c | 305d | 300f |
| Quantitative low titer (μL) | 285c | 394d | 290f |
| Quantitative high titer (μL) | 405c | 220e | 140f |
| Titer rangea | |||
| Low titer | 1:4 to 1:64g | 1:1 to 1:16 | 1:1 to 1:64 |
| High titer | 1:128 to 1:2,048 | 1:16 to 1:256 | 1:128 to 1:2,048 |
Derived from product instructions for use. Refer to respective manufacturer’s instructions for more information.
Reported based on participating public health laboratories’ inputs.
Includes 230 μL of dead volume for BioPlex Syphilis Total & RPR.
Includes 300 μL of dead volume for qualitative/screening and quantitative low titer for AIX 1000.
Includes 200 μL of dead volume for quantitative high titer specimen for AIX 1000.
Includes 190 μL of dead volume for ASI Evolution.
A reactive specimen with titer less than 1:4 is reported as <1:4, and for greater than 1:64 as >1:64, requiring further testing following instructions from the manufacturer’s package insert.