Table 4.

Performance of whole blood measurement by SH-SAW biosensor.

Test Item Marker
CRP Lp(a) apoB SARS-CoV-2 S Protein
Detection
Method
Pre-mix with
secondary antibody
Directly measurement Dilution measurement Directly measurement
Test time 3 min 3 min 30 s 40 s
Linearity range 1.9–118 μg/mL 83–1402 μg/mL 51–2022 μg/mL 50–1500 BAU/mL
Sensitivity LoQ = 1000 ng/mL LoD = 390 ng/mL LoB = 170 ng/mL LoD = 58.3 g/mL
LoB = 14.5 g/mL
LoD = 80.1 g/mL
LoB = 54.7 g/mL
LoD = 41.91 BAU/mL LoB = 27.48 BAU/mL
Comparison R = 0.986/
Slope = 1.003 (n = 170) *
R = 0.9698/
Slope = 1.0598 (n = 55) #
R = 0.9287/
Slope = 0.9741 (n = 55) %
NPA = 98.7% (n = 79)/PPA = 94.3% (n = 35) @

* Compared with Beckman Coulter CRP Latex; # Compared with Abbot Lp(a) Reagent Kit; % Compared with Advia Chemistry—Apolipoprotein B; @ Compared with cPass™ SARS-CoV-2 Neutralization Antibody Detection Kit; Positive percentage agreement (PPA); Negative percentage agreement (NPA); Limit of quantification (LoQ); limit of blank (LoB); and limit of detection (LoD).