Table 3.
Quantification of BCR::ABL1 in CML samples using DualQuant and Mobius commercial kit. Samples with BCR::ABL1 detection but lower than the quantification limit are represented by “Below LoQ”.
DualQuant | Mobius | |||||
---|---|---|---|---|---|---|
Samples | BCR::ABL1 Ct | ABL1 Ct | % | BCR::ABL1 Ct | ABL1 Ct | % |
1 | 28.79 | 21.87 | 0.074 | 31.52 | 23.11 | 0.085 |
2 | 31.44 | 24.14 | 0.066 | 34.69 | 24.83 | Below LoQ |
3 | 32.37 | 24.28 | 0.040 | 35.28 | 24.98 | Below LoQ |
4 | 29.27 | 21.65 | 0.061 | 32.08 | 22.6 | 0.041 |
5 | 32.1 | 23.14 | 0.019 | 35.01 | 24.15 | Below LoQ |
6 | ND | 25.75 | NA | 37.97 | 26.59 | Below LoQ |
7 | 24.08 | 23.8 | 8.459 | 26.35 | 24.83 | 9.377 |
8 | ND | 21.77 | NA | ND | 22.67 | NA |
9 | ND | 23.8 | NA | ND | 24.88 | NA |
10 | 28.4 | 22.64 | 0.457 | 31.15 | 24.12 | 0.218 |
11 | 34.78 | 26.03 | Below LoQ | 36.63 | 24.13 | Below LoQ |
12 | 33.96 | 29.75 | Below LoQ | ND | 37.98 | NA |
13 | 30 | 25.28 | 0.714 | 32.55 | 26.05 | 0.309 |
14 | 28.16 | 24.46 | 1.365 | 29.17 | 24.56 | 1.136 |
15 | 31.39 | 23.74 | 0.141 | 32.91 | 24.07 | 0.063 |
16 | 28.84 | 22.66 | 0.117 | 32.35 | 23.64 | 0.069 |
17 | 31.27 | 22.75 | 0.029 | 34.87 | 23.85 | Below LoQ |
18 | ND | 23.35 | NA | ND | 24.37 | NA |
19 | 24.96 | 25.05 | 5.890 | 26.82 | 24.23 | 4.531 |
ND: Not detected. NA: not available.