. 2023 Jun 14;10:1193805. doi: 10.3389/fcvm.2023.1193805

Table 1.

Device programming and automatic remote monitoring alert setting.

Trigger type ms/bpm Detection
AF Low—12% RR variability
HVR 180 bpm
Bradycardia 40 bpm
SRD 40%
Asystole 3 s
Device Type of alarm
Programmer triggered message received Yellow warning
ERI Red warning + notification
Backup mode active Red warning + notification
Sensing
Ventricular sensing amplitude (daily mean): <0.05 mV Red warning + notification
Atr. and ven. arrhythmia
Mean ventricular rate during AF > 150 bpm for >5% of day Yellow warning
Number of asystole episodes: at least one Red warning + notification
Number of high ventricular rate episodes (HVR): at least one Red warning + notification
Number of bradycardia episodes: at least one Yellow warning + notification
Number of sudden rate drop episodes: at least one Yellow warning + notification
Episode
Episode details received Yellow warning
Remote monitoring
First message received Yellow warning
No messages received for 5 days Yellow warning + notification
HM follow-up transmission has arrived Yellow warning

AF, atrial fibrillation; HVR, high ventricular rate; SRD, sudden rate drop; ERI, elective replacement indicator; HM, home monitoring.