Table 3.
Bridging stent grafts and target vessels in n = 44 patients.
| Celiac trunk | Superior mesenteric artery | Right renal artery | Left renal artery | ∑ | ||
|---|---|---|---|---|---|---|
| Stent type (%) | Viabahn VBX | 11 (25) | 12 (27) | 15 (34) | 11 (25) | 49 (28) |
| Advanta V12 | 24 (55) | 24 (55) | 19 (43) | 23 (52) | 90 (51) | |
| iCover | 8 (18) | 8 (18) | 6 (14) | 6 (14) | 28 (16) | |
| Other | 1 (2) | — | 4 (9) | 2 (5) | 7 (4) | |
| Additional lining (uncovered) (%) | 3 (7) | 1 (2) | 5 (11) | 4 (9) | 13 (7) | |
| Bridging stent extension (%)b | 16 (36) | 8 (18) | 8 (18) | 8 (18) | 40 (23) | |
| Target success (%) | 44 (100) | 44 (100) | 44 (100) | 42 (96) | 178 (99) | |
| Vessel angulationa (°) | 44.1 ± 20.7 | 41.7 ± 14.2 | 64.6 ± 23.5 | 67.3 ± 17.8 | — | |
| Vessel diametera (mm) | 6.9 ± 1.4 | 8.2 ± 1.7 | 5.9 ± 0.9 | 5.9 ± 1.1 | — | |
Continuous data presented as mean ± standard deviation.
Additional stent implantation necessary.