. 2023 Jul;64(7):1131–1137. doi: 10.2967/jnumed.122.264455

TABLE 3.

Number of Ordered-Subset Expectation Maximization Updates in Individually Optimized Reconstructions

Parameter Gaussian
filter (mm)
No. of OSEM updates for maximized SSR ME240
(20 kBq/mL),
3-min scan
HE80
(95 kBq/mL)
HE240
(94 kBq/mL)
ME80
(90 kBq/mL)
ME240
(90 kBq/mL)
HE80
(30 kBq/mL)
HE240
(29 kBq/mL)
ME80
(29 kBq/mL)
ME240
(29 kBq/mL)
AC 9 75 90 120 180 42 48 180 90 12
12 900 120 240 720 42 90 900 300 12
16 900 600 900 600 120 120 900 300 16
SC+AC 9 90 42 42 75 34 75 120 360 12
12 120 30 42 90 20 90 120 60 12
16 180 34 120 120 24 180 120 42 12

OSEM = ordered-subsets expectation maximization; HE80 = HE collimator with x-ray window; HE240 = HE collimator with γ-ray window; ME80 = ME collimator with x-ray window; ME240 = ME collimator with γ-ray window.

Table shows optimal number of ordered-subset expectation maximization updates, according to SSR, for each acquisition, as well as 224Ra concentration in phantom at time of imaging. Scan time was 30 min where not otherwise stated.