Table 1.
Multi-stepwise gradient elution program and multiple reaction monitoring (MRM) transitions and their associated mass spectrometric parameters for analysis of target analytes.
Gradient elution program | MRM and mass spectrometric parameters | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
|
|
|||||||||
Time, min | (A), % | (B), % | Flow rate, mL/min | Analyte | RT, min | Ion type | Ion transition | CE, eV | Dwell time, ms | Fragmentor, V |
| ||||||||||
0.01 | 95 | 5 | 0.75 | 3OH-COT | 3.1 | Q | 193.1 → 80.0 | 28 | 175 | 115 |
3.00 | 85 | 15 | 0.75 | C | 193.1 → 58.1 | 36 | 175 | 115 | ||
5.00 | 70 | 30 | 0.75 | L | 196.1 → 80.1 | 28 | 175 | 115 | ||
6.00 | 15 | 85 | 0.75 | COT | 4.2 | Q | 177.1 → 80.1 | 20 | 175 | 115 |
8.00 | 15 | 85 | 0.75 | C | 177.1 → 98.0 | 20 | 175 | 115 | ||
9.00 | 0 | 100 | 1.00 | L | 180.1 → 80.1 | 28 | 175 | 120 | ||
11.00 | 0 | 100 | 1.00 | COOH-THC | 8.1 | Q | 343.4 → 299 | 17 | 175 | 204 |
12.00 | 95 | 5 | 1.00 | C | 343.4 → 245 | 29 | 175 | 204 | ||
13.00 | 95 | 5 | 1.00 | L | 352.5 → 308 | 21 | 175 | 204 |
Note: (A) = 6.5 mM ammonium acetate (pH 5.1), (B) = acetonitrile, RT = retention time, CE = collision energy, 3OH-COT = trans-3′-hydroxycotinine, COT = cotinine, COOH-THC = 11-nor-9-carboxy-Δ9-tetrahydrocannabinol, Q = quantification ion, C = confirmation ion, L = labeled ion. The mass spectrometer divert valve position was switched to waste during the following time segments: 0–2.6 min, 5.0–7.5 min, 8.5–13 min.