Table 1.

Validation of an open-canal automatized d-lactate method.

Criteria Quality indicator Result
Measurement range
Limit of Detection (LoD) Blank test 0.02 µmol.L−1
Limit of Quantification (LoQ) <10% relative standard deviation (RSD) 40 µmol.L−1
Superior limit of linearity Need for dilution at high-concentration >15 mmol L−1
Analytical performances
Within-run imprecision Coefficient of Variation (CV%) IQClow: 1.98%
IQCmedium: 1.24%
IQChigh: 0.87%
Between-run imprecision IQCmedium: 2.21%
Robustness CV ratio 1.63
Interferences
Hemolysis ±10% maximal deviation 300 mg dL−1
Icterus 103 µmol.L−1
Lipemia L index of 47 (highly opalescent)
l-Lactate Dehydrogenase 1000 IU.L−1
Storage time 72 hours in plasma at 4 °C