Robustness expressed in % mean recovery and % relative standard deviation for GEM and BET

Parameter Mean of area % Mean recovery Standard deviation %RSD
GEM
 Mobile phase (75:25) 1,47,425 99.61 0.015 0.252
 Mobile phase (80:20) 1,48,821 100.55 0.053 0.910
 Mobile phase (85:15) 1,47,800 99.86 0.025 0.427
 Flow rate (0.9 ml/min) 1,49,147 100.77 0.062 1.064
 Flow rate (1.0 ml/min) 1,50,343 100.99 0.071 1.219
 Flow rate (1.1 ml/min) 1,47,460 99.64 0.016 0.268
 Column Temp. (20 °C) 1,49,229 100.83 0.064 1.103
 Column Temp. (25 °C) 1,49,715 100.59 0.055 0.942
 Column Temp. (30 °C) 1,50,691 101.82 0.104 1.796
BET
 Mobile phase (75:25) 61,101 101.35 0.058 1.040
 Mobile phase (80:20) 60,583 100.48 0.023 0.421
 Mobile phase (85:15) 60,767 100.79 0.036 0.644
 Flow rate (0.9 ml/min) 61,266 99.51 0.014 0.259
 Flow rate (1.0 ml/min) 62,144 100.93 0.041 0.745
 Flow rate (1.1 ml/min) 61,599 100.05 0.007 0.121
 Column Temp. (20 °C) 61,180 101.48 0.063 1.133
 Column Temp. (25 °C) 63,467 100.95 0.042 0.757
 Column Temp. (30 °C) 60,699 100.68 0.031 0.564