Table 2.
Model | Noise | SNR | PPV | NPV | Specificity | Sensitivity | AUROC (95% CI) | AUPRC |
---|---|---|---|---|---|---|---|---|
Noise-adapted | None | N/A | 0.203 | 0.990 | 0.676 | 0.924 | 0.896 (0.886–0.905) | 0.455 |
Portable ECG | 0.5 | 0.152 | 0.993 | 0.523 | 0.956 | 0.871 (0.861–0.882) | 0.392 | |
Portable ECG | 2 | 0.188 | 0.992 | 0.636 | 0.940 | 0.889 (0.880–0.899) | 0.439 | |
Electrode motion | 0.5 | 0.126 | 0.993 | 0.395 | 0.971 | 0.858 (0.846–0.869) | 0.361 | |
Electrode motion | 2 | 0.167 | 0.991 | 0.579 | 0.945 | 0.885 (0.875–0.895) | 0.426 | |
Muscle artifact | 0.5 | 0.175 | 0.989 | 0.608 | 0.927 | 0.871 (0.861–0.882) | 0.389 | |
Muscle artifact | 2 | 0.194 | 0.990 | 0.654 | 0.930 | 0.891 (0.881–0.900) | 0.449 | |
Baseline wander | 0.5 | 0.186 | 0.990 | 0.634 | 0.932 | 0.883 (0.873–0.893) | 0.427 | |
Baseline wander | 2 | 0.201 | 0.991 | 0.669 | 0.929 | 0.892 (0.883–0.902) | 0.457 | |
Standard | None | N/A | 0.207 | 0.988 | 0.688 | 0.910 | 0.895 (0.884–0.905) | 0.475 |
Portable ECG | 0.5 | 0.091 | 0.981 | 0.132 | 0.972 | 0.723 (0.706–0.739) | 0.200 | |
Portable ECG | 2 | 0.125 | 0.991 | 0.398 | 0.958 | 0.834 (0.822–0.847) | 0.329 | |
Electrode motion | 0.5 | 0.088 | 0.988 | 0.079 | 0.990 | 0.792 (0.779–0.806) | 0.239 | |
Electrode motion | 2 | 0.152 | 0.988 | 0.537 | 0.925 | 0.855 (0.843–0.866) | 0.333 | |
Muscle artifact | 0.5 | 0.100 | 0.991 | 0.210 | 0.979 | 0.807 (0.795–0.820) | 0.245 | |
Muscle artifact | 2 | 0.181 | 0.988 | 0.631 | 0.912 | 0.864 (0.853–0.875) | 0.344 | |
Baseline wander | 0.5 | 0.095 | 0.993 | 0.158 | 0.987 | 0.802 (0.789–0.816) | 0.236 | |
Baseline wander | 2 | 0.169 | 0.986 | 0.599 | 0.908 | 0.855 (0.843–0.866) | 0.322 |
SNR signal-to-noise ratio, PPV positive predictive value, NPV negative predictive value, AUROC area under receiver operating characteristic curve, CI confidence interval, AUPRC area under precision-recall curve.