Table 2. Relative Error and Mean of the Replicates Experimentally Obtained Using Both the Approach Developed and the Regular sp Approach over 8 Months for Five Different AuNPs (n = 7) and One AgNP (n = 5, Last Row)a.
diameter TEMb | this approach |
regular sp |
||
---|---|---|---|---|
diameterc | error (%) | diameterc | error (%) | |
16.4 ± 0.3 | 16.6 ± 1.2 | 1.2 | 16.3 ± 1.2 | 0.61 |
24.4 ± 0.3 | 24.7 ± 1.6 | 1.1 | 23.2 ± 1.5 | 4.9 |
27.4 ± 0.5 | 27.1 ± 1.7 | 1.0 | 25.5 ± 1.7 | 6.9 |
39.6 ± 0.7 | 39.5 ± 2.8 | 0.24 | 41.0 ± 3.1 | 3.5 |
73.3 ± 1.3 | 70.0 ± 7.4 | 4.5 | 61.9 ± 7.6 | 16 |
(68.1 ± 8.0)d | (7.1)d | |||
40.1 ± 0.6 (AgNP) | 39.5 ± 3.1 | 1.5 | 51.7 ± 3.2 | 29 |
(39.3 ± 5.1)d | (2.0)d |
The reference standard (AuNP, 18.7 nm) was measured once at the beginning of the time period and every working day for the developed and the regular sp approach, respectively.
Uncertainty corresponds to the standard error of the mean (k = 2) of the TEM images.
Combined standard uncertainty comprising individual uncertainty and reproducibility of the n replicates (95% confidence interval).
Results obtained when using AuNPs of 39.6 nm for the AuNPs of 73.3 nm and AgNPs of 51.3 nm for the AgNPs of 40.1 nm as reference standards, respectively.