Table 2. Relative Error and Mean of the Replicates Experimentally Obtained Using Both the Approach Developed and the Regular sp Approach over 8 Months for Five Different AuNPs (n = 7) and One AgNP (n = 5, Last Row)a.

diameter TEMb this approach
regular sp
diameterc error (%) diameterc error (%)
16.4 ± 0.3 16.6 ± 1.2 1.2 16.3 ± 1.2 0.61
24.4 ± 0.3 24.7 ± 1.6 1.1 23.2 ± 1.5 4.9
27.4 ± 0.5 27.1 ± 1.7 1.0 25.5 ± 1.7 6.9
39.6 ± 0.7 39.5 ± 2.8 0.24 41.0 ± 3.1 3.5
73.3 ± 1.3 70.0 ± 7.4 4.5 61.9 ± 7.6 16
(68.1 ± 8.0)d (7.1)d
40.1 ± 0.6 (AgNP) 39.5 ± 3.1 1.5 51.7 ± 3.2 29
(39.3 ± 5.1)d (2.0)d
a

The reference standard (AuNP, 18.7 nm) was measured once at the beginning of the time period and every working day for the developed and the regular sp approach, respectively.

b

Uncertainty corresponds to the standard error of the mean (k = 2) of the TEM images.

c

Combined standard uncertainty comprising individual uncertainty and reproducibility of the n replicates (95% confidence interval).

d

Results obtained when using AuNPs of 39.6 nm for the AuNPs of 73.3 nm and AgNPs of 51.3 nm for the AgNPs of 40.1 nm as reference standards, respectively.