. 2023 Jul 10;12(7):11. doi: 10.1167/tvst.12.7.11

Table.

Coefficient of Repeatability (CoR) of the Microperimetry Visual Sensitivity Measures for Sequential Pairwise Comparisons Across the Three Tests

Comparison Between
Outcome Measure Test 1 vs. 2 Test 2 vs. 3 Difference (%)a P b
Mean sensitivity (95% CI) 2.5 (2.1–2.8) 1.4 (1.0–1.8) 79 <0.001
Pointwise sensitivity (95% CI)
 Overall 4.7 (4.3–5.0) 3.9 (3.5–4.2) 21 <0.001
 <6 dB 3.8 (3.2–4.2) 2.2 (1.6–2.8) 73 <0.001
 6 to 12 dB 12.9 (12.1–13.8) 8.2 (7.3–9.1) 57 <0.001
 12 to 18 dB 8.9 (8.2–9.5) 7.6 (6.9–8.2) 17 0.008
 18 to 24 dB 4.9 (4.5–5.3) 4.2 (3.8–4.6) 17 0.031
 >24 dB 3.0 (2.6–3.4) 2.8 (2.5–3.3) 7 0.567
a

Difference between the first and second pairs of tests.

b

Comparison of the CoR values between the two sequential pairs of tests.