Table 5.
Measurement Error in Naturally Ventilated Fine-Wire Thermocouple Sensors.
Aspirated Ref. → | Thermocouple Yankee 2010 | PRT MetOne 327-C | ||
---|---|---|---|---|
Test shield ↓ | Error (µ ± σ; °C) | Shapiro–Wilk | Error (µ ± σ; °C) | Shapiro–Wilk |
Gill (N = 2072) | 0.11 ± 0.34 | 0.696, p < 0.001 | 0.12 ± 0.45 | 0.968, p < 0.001 |
Custom (N = 691) | 0.11 ± 0.40 | 0.701, p < 0.001 | 0.12 ± 0.49 | 0.945, p < 0.001 |
Data are from the Savannah River National Laboratory calibration experiment [171].