Skip to main content
. 2022 Apr 12;17(4):951–958. doi: 10.1177/19322968221088608

Table 4.

Comparison of HCL System Pivotal Trial Outcomes and Real-World HCL System Use Analyses.

HCL system pivotal trial a
(N = 124)
First real-world HCL system use analysis a
(n = 3141)
Present study real-world HCL system use analysis
(n = 19 929)
Before Auto Mode initiation (Run-in) After Auto Mode initiation (Study) P b Before Auto Mode initiation 3 months after Auto Mode initiation P b Before Auto Mode initiation 3 months after Auto Mode initiation P b
Auto Mode, % 82.2 80.8 89.4
SG, mg/dL 150 151 .267 159 152 <.001 167 158 <.001
 <54 mg/dL 1.3 0.8 <.001 0.6 0.5 <.001 0.5 0.6 <.017
 <70 mg/dL 5.5 3.0 <.001 2.7 2.1 <.001 2.1 2.1 .117
 70-180 mg/dL 67.1 72.4 <.001 66.0 73.3 <.001 62.4 69.3 <.001
 >180 mg/dL 27.4 24.5 <.001 31.4 24.6 <.001 35.6 28.7 <.001
 >250 mg/dL 6.9 5.6 .007 8.1 5.4 <.001 10.2 7.2 <.001
TDD, units 47.5 50.9 <.001 49.8 51.9 <.001 50.1 51.1 <.001

Data are shown as mean. The pivotal trial Run-in period (before Auto Mode initiation) was ~2 weeks and the Study period was 3 months. The mean percentage of time in Auto Mode, sensor glucose, percentage of time in sensor glucose ranges, and total daily insulin dose delivered before and 3 months after Auto Mode initiation are shown for the HCL system pivotal trial, the first real-world HCL system use analysis and the present real-world HCL system use analysis.

Abbreviations: HCL, hybrid closed-loop; SG, sensor glucose; TDD, total daily insulin dose.

a

As reported in Stone et al. 9

b

Comparison of before versus after Auto Mode initiation.