Skip to main content
. 2023 Jul 10;34(2):142–152.

Table 1.

Evaluation of HSGC procedure with 20% TEa (CLIA2019)

Concentration utrg sr sr/√(nal) (A) utrg/3 (B) Conclusion
Low 5.60 0.87 0.02 1.87 Good precision1
Medium 13.4 0.61 0.02 4.47
High 26.8 1.31 0.03 8.93

1 criterion (A) < (B) conclusion indicates ‘good precision’

Abbreviations: utrg represents the target uncertainty; sr is the standard deviation; and nal denotes the unit for measurement.