Risk assessment plan for non-integrated POCT in analytical testing phase

Risk Assessment Components Possible sources of error Risk Mitigation: Manage/Prevent Errors from Occurring
Testing Personnels’ Training and Competency Not trained POCT user
Not competent POCT user
Improper training or expired competency records
Proper POCT technique/ system maintenance /QC not being accurately followed
  • Central laboratory to provide proper initial training and competency

  • Competency assessment frequency to be decided based on risk assessment

  • Compliance monitoring

  • Follow established policies and procedures

  • Testing personnel training and competency records

  • Traceability of POCT users

Test system- Calibration Calibration omitted
Calibration out and patient results reported
There is no mechanism, such as an operator lockout, to ensure only trained personnel use the test system
  • Ensure POCT users follow standard operating procedures through trainings, competency assessment, audits

  • Test process flow charts

  • POCT user manual/electronically shared QC and calibration records with POCT coordinator

Test system- Maintenance Maintenance procedures not consistent with manufacturer’s instructions
  • Train POCT users for POCT device maintenance

Test system-Reagent Expired reagents
Wrong lot of reagents
  • Train POCT users to check reagent expiry prior to use

  • Manufacturer’s instructions to be followed for reagents storage

Test system- QC Expired QC
Wrong lot of QC
Unable to interpret QC
Unable to trouble shoot if QC is out
  • Prevent QC degradation during storage and use

  • Training of POCT users on interpreting QC before release of patient results

  • Maintain QC logs (manual if electronic not possible)

  • Assure POCT user review each QC before patient result release – through trainings

  • Training of POCT users on troubleshooting

Environment-room Temperature Room temperature not appropriate
  • Daily room temperature logs

  • POCT coordinator to sign off logs periodically

Environment- Storage conditions Sample / reagents/ cartridge/ strips storage conditions not appropriate
  • Daily refrigerator temperature logs where POCT items are stored.

  • POCT coordinator to sign off logs periodically