Table 1.

Regression and validation data for the determination of SER.DMP and DMP by the proposed methods

Drug Name SER.DMP
Parameters HPLC method TLC densitometric method
Wavelength (nm) 220 nm 220 nm
Linearity range 2.5— 25 (μg/mL) 5— 25 (μg/spot)
LOD 0.051(μg/mL) 0.184 (μg/spot)
LOQ 0.165 (μg/mL) 0.202 (μg/spot)
Regression Equation y a = 0.0989 C b + 0.0331 y a = 0.0824 C b + 0.0216
 Correlation coefficient (r) 0.9998 0.9997
 Accuracy (% R) ± SD 100.43 ± 1.02 99.68 ± 1.03
Precision (% RSD)
 Repeatability c 1.021 1.025
 Intermediate precision d 1.015 1.039
Robustness (% RSD)
 Mobile phase contents ratio (± 2%) 0.812 1.386
 Detection wavelengths (± 2 nm) 1.349 1.526
 Flow rate (± 0.1 ml/min.) 1.778 ـــــــ
Drug Name DMP
Parameters HPLC method TLC densitometric method
Wavelength (nm) 220 nm 220 nm
Linearity range 2.5–25 (μg/mL) 5–25 (μg/spot)
LOD 0.098 (μg/mL) 0.115 (μg/spot)
LOQ 0.186 (μg/mL) 0.237 (μg/spot)
Regression Equation y a = 0.0897 C b + 0.0522 y a = 0.0905 C b + 0.0612
 Correlation coefficient (r) 0.9997 0.9996
 Accuracy (% R) ± SD 100.15 ± 1.1 99.50 ± 1.2
Precision (% RSD)
 Repeatability c 1.105 1.160
 Intermediate precision d 1.251 1.249
Robustness (% RSD)
 Mobile phase contents ratio (± 2%) 0.812 1.128
 Detection wavelengths (± 2 nm) 1.349 1.630
 Flow rate (± 0.1 ml/min.) 1.778 ـــــــ

a Peak area of mixture of SER.DMP and DMP

b Concentration in μg/ml for HPLC method and in μg/spot for TLC densitometric method

c The intraday (n = 3), average of three concentrations of SER.DMP and DMP (10, 15 and

20 μg/mL) for HPLC method and (10,15 and 20 μg/spot) for TLC densitometric method

repeated three times within the day

d The interday (n = 3), average of three concentrations of SER.DMP and DMP (10,15

and 20 μg/mL) for HPLC method and (10,15 and 20 μg/spot) for TLC densitometric

method repeated three times in three days