Skip to main content
. 2023 Jun 14;40(7):1677–1687. doi: 10.1007/s10815-023-02828-6

Table 4.

Measurement of trueness in the two labs. Bias was calculated using inter-laboratory comparison (ILC) according to the following formula: Bias (%) = ((mv)/v) × 100; m is the mean of values measured by the lab for the same batch of ILC and v is the target value provided by the quality control body

ILC 1 ILC 2 ILC 3
Bias (%) lab 1 −0.07 −0.02 −0.03
Bias (%) lab 2 −0.02 −0.03 0.03