Skip to main content
. 2023 Jul 5;8(28):25101–25113. doi: 10.1021/acsomega.3c01791

Table 3. Robustness Result of EXE and GEN.

chromatographic parameter Modification theoretical plates retention time tailing factor
Robustness Data of EXE
wavelength (nm) 242 7494 2.10 1.47
  245 7540 2.12 1.48
  247 7646 2.11 1.45
flow rate (mL/min) 1.0 7574 2.10 1.42
  0.8 7454 2.14 1.45
  1.2 7604 2.18 1.48
composition of mobile phase 30:70 7543 2.11 1.41
  20:80 7555 2.13 1.42
  40:60 7621 2.15 1.45
Robustness Data of GEN
wavelength (nm) 268 6678 1.66 1.54
  270 6687 1.67 1.56
  271 6798 1.68 1.57
flow rate (mL/min) 1.0 6680 1.65 1.54
  0.8 6779 1.67 1.55
  1.2 6987 1.69 1.58
composition of mobile phase 30:70 6789 1.64 1.56
  20:80 6877 1.67 1.58
  40:60 6988 1.69 1.59