Skip to main content
. 2023 Jul 5;8(28):25101–25113. doi: 10.1021/acsomega.3c01791

Table 4. Accuracy Results of EXE and GEN.

sample name level actual concentration of sample (μg/mL) concentration of sample spiked (μg/mL) concentration of sample obtained (μg/mL) % recovery SD % RSD
EXE
EXE 80% 15 27 26.90 99.62 0.745 0.74
80% 15 27 27.30 101.11
80% 15 27 27.10 100.37
100% 25 50 50.10 100.21 0.995 0.96
100% 25 50 49.60 99.25
100% 25 50 49.15 98.30
120% 35 77 77.20 100.25 0.819 0.82
120% 35 77 77.37 100.48
120% 35 77 76.20 98.96
GEN
GEN 80% 15 27 27.29 101.07 0.551 0.55
  80% 15 27 26.99 99.96    
  80% 15 27 27.12 100.44    
  100% 25 50 49.50 99.00 1.216 1.21
  100% 25 50 50.66 101.32    
  100% 25 50 49.78 99.56    
  120% 35 77 77.70 100.90 0.775 0.77
  120% 35 77 77.15 100.19    
  120% 35 77 76.50 99.35