Baseline, procedural, and follow-up data

Aneurysm/side Baseline aneurysm dimensions Embolization coil volumes and packing densities Raymond–Roy*
NW W H V TX-PE V TX-E V TX-PE pack TX-E pack Total V Total pack TX-PE Total pack TX-E TX-E % total Stent assisted coiling PP 6 months 12 months
mm mm mm mm3 mm3 mm3 % % mm3 % % %
C6 seg ICA R 8.2 14.4 14.9 1618 120 460 28 0 460 7 28 100 Yes 3 3 2
C6 seg ICA L 2.0 6.0 8.3 156 20 76 49 15 91 22 58 84 2 1† 1
C6 seg ICA‡ L 5.5 7.8 4.5 143 10 38 26 7 45 12 31 85 Yes 3 1 1
C6 seg ICA R 2.7 6.2 5.7 115 12 47 41 13 60 22 53 78 1 1† 1
C6 seg ICA L 4.1 6.1 6.0 117 10 36 31 12 48 19 41 75 2 1
C6 seg ICA¶ R 3.1 6.1 4.8 94 8 31 34 10 42 20 45 75 1 1 1
C6 seg ICA L 3.4 4.8 5.2 63 6 24 38 5 28 18 45 83 2 2
PCom A** R 4.7 8.7 8.1 321 17 65 20 15 80 10 25 82 Yes 2 2† 2
ACA A1–A2 L 4.0 6.8 5.0 121 16 62 51 15 77 26 64 80 Yes 2 1 1

Aneurysm volumes, coil volumes, and packing densities were determined using the calculator at www.angiocalc.com in March 2022. Online supplemental table 2 contains specifics on all the coils used in each case.

*Determined by fluoroscopy guided catheter angiography at the time points indicated, with some exceptions (see below).

†Determined by MR angiography at 10 months because of COVID-19 pandemic restrictions to clinical operations.

‡Illustrated in figure 3.

§Determined by catheter angiography at 19 months because of COVID-19 pandemic restrictions to clinical operations.

¶Illustrated in figure 2.

**Illustrated in figure 4.

ACA A1-A2, anterior cerebral artery A1–A2 segment; C6 seg ICA, ophthalmic segment of the internal carotid artery; H, height; NW, neck width; PCom A, posterior communicating artery; PP, at case completion angiography; Raymond–Roy, Raymond–Roy occlusion classification; Total pack TX-E, packing density based on the total volume of all coils used in the aneurysm with the TrelliX embolic coil volume with expanded shape memory polymer; Total pack TX-PE, packing density based on the total volume of all coils used in the aneurysm with the TrelliX embolic coil volume prior to expansion of the shape memory polymer; Total V, total coil volume, including coils other than TrelliX embolic coils; TX-E pack, packing density based on the TrelliX embolic coil volume with expanded shape memory polymer; TX-E % total, proportion of the total coil volume that is TrelliX embolic coil volume, based on expanded shape memory polymer; TX-E V, TrelliX embolic coil volume with expanded shape memory polymer; TX-PE pack, packing density based on the TrelliX embolic coil volume prior to the expansion of the shape memory polymer; TX-PE V, TrelliX embolic coil volume prior to the expansion of the shape memory polymer; V, volume; W, width.