Liquid chromatography and mass spectrometry selected parameters.

Mobile phase Binary system of 0.1% Formic acid in H2O (A) and ACN (B) ESI source Positive ESI
High purity N2 gas Drying gas at 10 L/min with pressure (60 psi)
0.5 mL/min.
Elution time: 55 min.
Agilent Zorbax eclipse plus C18Column Length 150 mm ESI temperature: 350 °C
ID 4.6 mm Capillary voltage: 4000 V
Particle size 3.5 μm Modes Mass scan and MS2
T 22 ± 1 °C Collision gas High purity N2
Gradient elution system Time in min. % ACN Analytes SPB and its reactive metabolites
0 5 Mass parameters Fragmentor voltage (FV): 145 V
Amplitude: 1.25 V
25 75
40 75
55 5