Skip to main content
. 2023 Jul 7;14:1194848. doi: 10.3389/fmicb.2023.1194848

Table 1.

Performance of selected scintillation cocktails at various measurement times.

[CPM2] 1  min 2  min 10  min 60  min
ROTISZINT 23.48 ± 5.38 21.59 ± 4.28 20.60 ± 2.30 19.32 ± 0.99
AquaLight+ 19.70 ± 4.56 18.07 ± 3.33 19.75 ± 1.42 20.04 ± 0.64
Ultima Gold XR 23.07 ± 77.95 14.58 ± 2.87 15.96 ± 1.60 14.94 ± 0.63

Performance is presented by CPM2 (events counted by two of the three PMTs) of the counter background values and respective standard deviation.