Performance of selected scintillation cocktails at various measurement times.

[CPM3] 1  min 2  min 10  min 60  min
ROTISZINT 13.49 ± 3.56 (±3.67) 12.49 ± 2.63 (± 2.50) 12.62 ± 1.14 (±1.12) 12.52 ± 0.49 (±0.45)
AquaLight+ 12.92 ± 3.76 (±3.59) 11.82 ± 2.42 (±2.43) 12.97 ± 1.17 (±1.14) 13.32 ± 0.55 (±0.47)
Ultima Gold XR 10.31 ± 36.25 (±3.21) 8.03 ± 1.97 (±2.00) 7.80 ± 0.91 (±0.88) 7.90 ± 0.42 (±0.36)

Performance is presented by CPM3 (events counted by all three PMTs) of the counter background values and respective standard deviation. Values in parentheses are calculated values based on the Poisson distrubution. Except for the 1-min measurement of Ultima Gold XR, the calculated and measured values are in good agreement.