Skip to main content
. 2023 Jul 21;23:131. doi: 10.1186/s12911-023-02215-2

Table 6.

Comparison of our results with the Association for the Advancement of Medical Instrumentation (AAMI) standard

No. of Subjects MAE (mmHg) ME (mmHg) STD (mmHg)
AAMI Standard 30  > 85 /  < 5  < 8
MLM-Transformer ASBP 347 4.97 0.043 4.72
ADBP 347 2.99 0.026 2.39
MLM-Transformer w/personalization ASBP 347 2.41 0.037 2.72
ADBP 347 1.31 0.029 1.77