Skip to main content
. 2023 Jul 10;14:1204649. doi: 10.3389/fphar.2023.1204649

TABLE 1.

Evaluation of the Intra- and Inter-day precision, recovery and matrix effects by the proposed UPLC-MS/MS method for determination of macitentan and ACT-132577 in rat plasma (n = 6).

Analytes Preparation concentration (ng/mL Intra-day Inter-day Recovery (%) Matrix effects (%)
Precision RSD (%) Accuracy RE (%) Precision RSD (%) Accuracy RE (%)
Macitentan 4 6.90 10.05 9.36 8.62 86.43 86.16
160 5.24 6.57 2.79 7.59 95.94 92.87
1,600 4.79 9.55 3.67 7.77 90.91 91.54
ACT-132577 30 7.01 11.52 6.83 8.74 88.12 89.77
900 4.53 8.49 5.9 5.83 90.40 94.02
9000 3.68 2.28 2.49 2.92 92.18 93.18

Abbreviations: RSD, relative standard division; RE, relative error.