Table 3.
Troubleshooting of the UHPLC-EISA-TOF method for the quantification of DL-2-HG in biological samples
| Step | Problem | Observation | Possible reason | Solution |
|---|---|---|---|---|
| Derivatization (Step 16) | Derivatization did not work | Standards or samples do not turn into yellow or brownish color | Not all water was evaporated. Alternatively, it could also be possible that the sample is not very concentrated and thus the color is not that noticeable. | Make sure all water has been evaporated before adding L-DATAN solution. |
| Liquid chromatography instrument (Steps 24–26) | High column backpressure | Pressure near 400 bar and/or low reproducibility of the retention time | Column may be clogged | Clean the column using 2-propanol. If the problem persists, replace the column |
| Pressure fluctuation | The pressure is not constant | Valve or pump seal leaks and might be damaged | Replace the check valve and the pump seals | |
| Loss of chiral separation | The resolution of the peaks is lower than expected | Mobile phase A and B are not prepared correctly or column is defective | Make sure mobile phase A and B are well prepared and/or replace the column | |
| Mass spectrometry instrument (Steps 21–23) | No peaks appear | Ions of the compound are not observed | Some parameters are not set correctly setup and/or the derivatization step was not successful | Check the MS parameters and/or make sure the derivatization was conducted in the proper way |
| Poor mass accuracy | Peaks are not observed when they are extracted with a width of m/z ±0.01 | The system is not well calibrated | Calibrate the MS detector according to the manufacturer’s instructions | |
| Loss of sensitivity | Ion counts are too low | The ion source may be contaminated | Clean the ion source, spray shield and the glass capillary following the manufacturer’s instructions |