Table 4.

Results of assessing the trueness, precision and recovery of the different biological samples by the developed EISA-TOF method

K562 cells Urine FCS Serum

L-2-HG D-2-HG L-2-HG D-2-HG L-2-HG D-2-HG L-2-HG D-2-HG
Trueness % Mean recovery ± SD

10 μmol/L 97 ± 4 90 ± 5 98 ± 8 105 ± 5 105 ± 8 101± 4 103 ± 2 106 ± 6
50 μmol/L 95 ± 2 87 ± 2 100 ± 8 113 ± 10 100 ± 5 96.7 ± 0.3 106 ± 3 105 ± 5
70 μmol/L 97 ± 2 83 ± 5 98 ± 3 105 ± 5 99 ± 2 96 ± 3 105 ± 1 105 ± 2
Precision RSD (%)

Between-injection repeatability (n=3)a 5.0 5.0 3.6 3.9 1.1 2.2 1.3 2.2
Measurement repeatability (n=3)b 4.8 4.8 1.8 2.4 5.3 6.8 3.5 3.1
Intermediate measurement precision (n=9)c 4.9 4.9 4.6 4.4 6.7 7.3 3.8 5.2
Recovery (before and after the extraction) % Mean recovery ± SD

104 ± 8 104 ± 8 99 ± 2 94 ± 5 99 ± 3 94 ± 7 95 ± 3 99 ± 5
a

Three consecutive injections of the corresponding biological sample spiked with 10 μmol/L of L-2-HG, 10 μmol/L of D-2-HG, 10 μmol/L of L-2-HG-d3 and 10 μmol/L of D-2-HG-d3.

b

Three replicates of the corresponding biological sample spiked with 10 μmol/L of L-2-HG, 10 μmol/L of D-2-HG, 10 μmol/L of L-2-HG-d3 and 10 μmol/L of D-2-HG-d3 injected in triplicate on the same day.

c

Three replicates of the corresponding biological sample spiked with 10 μmol/L of L-2-HG, 10 μmol/L of D-2-HG, 10 μmol/L of L-2-HG-d3 and 10 μmol/L of D-2-HG-d3 injected during three consecutive days.