MRA improves sensitivity than MRI for the articular-sided partial-thickness rotator cuff tears

Lianxin Li 1,*, Jinlei Dong 1,*, Qinghu Li 1, Jinye Dong 2, Bingzhi Wang 1, Dongsheng Zhou 1, Fanxiao Liu 1,
PMCID: PMC10364937  PMID: 34939473

Abstract

Background:

This study aims to compare the diagnostic accuracy of magnetic resonance imaging (MRI) and MR arthrography (MRA) for the articular-sided partial-thickness rotator cuff tear (PTRCT).

Methods:

Three electronic databases, PubMed/Medline, Embase and Cochrane Library, were utilized to retrieve articles comparing the diagnostic value of MRA and MRI for detecting articular-sided PTRCTs. The pooled statistical indexes included sensitivity, specificity, positive/negative predictive value, diagnostic odds ratio (DOR) and the area under the receiver operating characteristic curve (AUC).

Results:

Eleven studies involving 1703 patients and 1704 shoulders were included. The pooled sensitivity, specificity, DOR and AUC and their 95% CIs of MRA to diagnose articular-sided PTRCTs were 0.81 (95% CI, 0.65–0.90), 0.96 (95% CI, 0.91–0.98), 68.14 (95% CI, 33.20–139.84) and 0.96 (95% CI, 0.94–0.97), respectively. The pooled sensitivity, specificity, DOR and AUC and their 95% CIs of MRI were 0.78 (95% CI, 0.65–0.87) and 0.97 (95% CI, 0.84–0.99), 47.82 (95% CI, 8.29–275.89) and 0.89 (95% CI, 0.86–0.92), respectively.

Conclusions:

This meta-analysis reveals that MRA has a better diagnostic value than that of MRI for the diagnosis of articular-sided partial-thickness rotator cuff tears because of an improvement of sensitivity.

Keywords: Rotator cuff tear, MRA, MRI, partial-thickness, shoulder pain

Introduction

Partial-thickness rotator cuff tear (PTRCT) was first described by Codman 1 and later divided into bursal-sided, articular-sided and intratendinous tears according to the depth and location of the tear. 2 It should be noted that articular-sided PTRCT was 2–3 times more common than bursal-sided tears.3,4

Numerous diagnostic studies focused on the evaluation of the performance of MRA, MRI or ultrasound (US) imaging for the diagnosis of full- or partial-thickness tears. A previous meta-analysis confirmed that three-dimensional US is very effective and highly accurate to detect full-thickness RCTs but may lack accuracy in the diagnosis of partial tears. 5 A recent network meta-analysis of diagnostic tests performed by our team reported that high-field MRA had the highest diagnostic value, then low-field MRA, followed by high-field MRI, high-frequency ultrasound, low-field MRI, and low-frequency ultrasound. 6 However, both of the meta-analyses were performed based on the full-, partial- or any tear. It is important to identify the subtype of partial-thickness tears.7,8 Furthermore, another meta-analysis also conducted by our team indicated that MRA and MRI have a similar diagnostic value for the diagnosis of bursal-sided partial-thickness rotator cuff tears. 9 However, there is no study focused on the articular-sided PTRCTs.

The meta-analysis aims to assess the diagnostic accuracy of MRA and MRI for detecting articular-sided PTRCTs based on all available scientific published data.

Materials and methods

This study was performed in accordance with the Preferred Reporting Items for a Systematic Review and Meta-analysis of Diagnostic Test Accuracy Studies (PRISMA-DTA) statement. 10

Data sources and search strategy

Two independent reviewers searched PubMed, Embase and the Cochrane Library (including Epub Ahead of Print) for titles from data inception to January 1, 2019. The vocabulary and syntax were specifically adapted according to the database. The combination of keywords and mesh terms were “diagnostic”, “diagnosis”, “rotator cuff”, “supraspinatus”, “infraspinatus”, or “subscapularis” and “MRI”, “magnetic resonance imaging”, “MRA” or “magnetic resonance arthrography”. The purpose, research question, and eligibility criteria for the search were determined a priori. Additionally, the reference lists of relevant articles and included studies were searched by hand for supplementary eligible records. Any disagreement was settled through the discussion of researchers until a consensus was reached.

Study screening

The included studies should confirm all criteria listed as follows: 1) study design, diagnostic accuracy study; 2) population, patients with suspected rotator cuff tear; 3) MRA/MRI test was performed; 4) the final diagnosis of articular-sided RCT was confirmed by predesigned gold standards and 5) diagnostic data (number of false/true-positive [FP/TP] and false/true-negative [FN/TN] cases) could be extracted or calculated to construct a two-by-two contingency table.

Data extraction

Following information was collected from all included articles into a pre-designed Microsoft Excel spreadsheet (Version 2010, Microsoft, Redmond, WA, USA) by two investigators blindly and repeatedly: the first author's surname, publication time of the original study, sources of origin, participant characteristics (number, age and gender), design of the study, gold standard, time from the diagnostic test to the gold standard, blinding, no. of readers, the experience of readers, clinical findings of patients, technical parameters of MRA/MRI including the administration of contrast agent (indirect or direct), vendor, model, magnetic field strength, various sequences, slice thickness, analyzed image plane. Data extraction from all included studies was completed in tandem by two independent reviewers. The spreadsheets were combined and each reviewer checked a random selection of the other's entries for quality control. Any discrepancies were resolved by consensus decision.

Quality assessment

Two investigators performed blindly the methodological quality of the included studies in duplicate using the Quality Assessment of Diagnostic Accuracy Studies-2 (QUADAS-2) tool. 11

Statistical analysis

Quantitative analyses in this study were conducted with the forest plots using STATA 12.0 Version (V. 12.0, StataCorp, College Station, TX). The whole process of searching, filtering, data extraction and quality assessment was implemented by two researchers (LFX & DJL) independently and repeatedly. For any discrepancy, a consensus was reached by discussion with an arbitrator (LLX). The pooled estimates indexes to assess the accuracy of MRA/MRI included sensitivity, specificity, positive likelihood ratio (PLR), negative likelihood ratio (NLR), diagnostic odds ratio (DOR) and the area under the receiver operating characteristic curve (AUC). Sensitivity is the proportion of true positive tests out of all patients with a condition. The formula is below: Sensitivity = (True Positives) / (True Positives + False Negatives). Specificity is the percentage of true negatives out of all subjects who do not have a disease or condition. The formula is below: Specificity = (True Negatives) / (True Negatives + False Positives); A positive likelihood ratio is the “probability that a positive test would be expected in a patient divided by the probability that a positive test would be expected in a patient without a disease.”; A negative likelihood ratio is “the probability of a patient testing negative who has a disease divided by the probability of a patient testing negative who does not have a disease”. The formulas for the likelihood ratios are given below: Positive Likelihood Ratio = Sensitivity / (1-Specificity); Negative Likelihood Ratio = (1- Sensitivity) / Specificity; The Diagnostic Odds Ratio (DOR) is defined as the ratio of the odds of the test being positive if the subject has a disease relative to the odds of the test being positive if the subject does not have the disease. DOR = Sensitivity  × Specificity / ((1−Sensitivity)  × (1−Specificity)) = PLR + /NLR; A receiver operating characteristic curve, or ROC curve, is a graphical plot that illustrates the diagnostic ability of a binary classifier system as its discrimination threshold is varied. When using normalized units, the area under the curve (often referred to as simply the AUC) is equal to the probability that a classifier will rank a randomly chosen positive instance higher than a randomly chosen negative one. The sensitivity and specificity of MRI and MRA for the detection of partial-thickness tears were calculated and a chi-square test was performed on this data. A p-value of < 0.05 was considered statistically significant. All analyses were performed with the SPSS program (version 17, IBM SPSS, USA). Heterogeneity among the included studies was assessed using the I 2 statistic. The meta-regression and subgroup analyses were conducted to explore the available source of heterogeneity. Publication bias was performed using Deeks' Funnel Plot Asymmetry Test. Additionally, we performed meta-regression analysis and the patient sample size (≥100 or <100), publication year (before or after 2014), magnetic field strength (3.0-T or not), number of readers (≥2 or 1), Blinding (Yes or No) and QUADAS-2 score (≥10 or <10), design of the study (prospective or retrospective) as well as muscle-tendon were used as covariates.

Results

Studies retrieved and characteristics

In total, 2192 articles were identified by searching three databases and removing duplicates. The screening of the reference lists of these and other relevant articles yielded four additional studies. After screening remaining titles and abstracts, and identifying related full-text, 11 studies8,1221 published during the period from 2007 to 2018 were included for quantitative analysis.

The sample size of the included studies ranged from 10 to 333 with a total of 1703 patients. The main characteristics and technical parameters of MRI and MRA in the included studies are presented in Table 1. According to the QUADAS-2 score, 413,17,20,21 (36.4%), 28,19 (18.2%), 31416 (27.3%), 1 12 (9.1%) and 1 (9.1%) 18 study scored 11, 10, 9, 8 and 7, respectively.

Table 1.

Main characteristics and technical parameters of the included studies.

Study, year Country Patients (Male, %) Mean age (years) Shoulders Test Sequence Analyzed image plane No. of readers Reader experience (years) Study design Inclusion interval Muscle tendon QUADAS- 2 score
Fritz, 2007 USA 230 (65.2) 43 238 1.5-T MRI FS, T1WI, GE transverse, cor obl, sag obl 2 NA R 04.2000–07.2004 SSP or ISP 8
Magee, 2009 USA 150 (69.0) 31 150 3.0-T MRA T1WI FSE, FS T2WI FSE Ax, cor obl, sag 2 10/10 R 01.2007–07.2007 SSP 11
3.0-T MRI T1WI FSE, FS T2WI FSE, FS T2WI FSE
Oh, 2009 Korea 36 (44.4) 53.9 36 3.0-T MRA FS T1WI FSE, T2WI FSE, 3D fast FS GRE Ax, cor obl, sag obl 2 NA P 03.2006–06.2006 SSP or ISP 9
Chun, 2010 Korea 202 (54.5) 51 202 1.5-T MRA FS T1WI SE,T2WI FSE transverse, sag obl, cor obl 2 > 10 R NA All 10
Choo, 2012 Korea 49 (44.9) 55.6 49 3.0-T MRA FS T1WI FSE, T2WI FSE, 3D FS T1WI FSE Ax, cor obl, sag obl 2 Yes R 08.2010-04.2011 All 9
Modi, 2013 UK 103 (73.8) 30 103 3.0-T MRA FS T1WI SE, STIR, fs T2WI FSE, fs T1WI SE Ax, cor obl, sag obl 3 Yes NA 11.2006-07.2011 All 9
Lee, 2014 Korea 205 (47.8) 56.5 206 3.0-T MRA FS T1WI FSE, 3D FS T1WI FSE Ax, cor obl, sag obl 1 8 R 03.2011-07.2012 SSP or ISP 11
Choo, 2015 Korea 231 (42.0) 59 231 3.0-T MRA T2WI FSE, FS T1WI FSE cor obl, sag obl 1 8 R 01.2011-12.2013 SSP or ISP 7
Lee, 2015 Korea 333 (48.0) 56.9 333 3.0-T MRA FS T1WI, T2WI FSE Ax, cor obl, sag obl 2 10/7 R 03.2011-09.2013 SSP or ISP 10
3.0-T MRI GE, FS FSE PDW, T2WI FSE
Lo, 2016 China 146 (65.1) 48.3 146 1.5-T MRI T1WI, FS T2WI, SE DWI cor obl 2 20/5 R 01.2012-07.2013 All 11
Perez, 2018 USA 10 (100) 16.7 10 1.5-T MRI FS T2, PD, T1WI Ax, cor obl, sag obl NA NA R 01.2010-10.2016 All 11

P prospective, R retrospective, PD proton-density, PDW proton-density-weighted, FSE fast spin-echo, STIR short Tau-inversion recovery, GE gradient echo, TSE turbo spin-echo, FS fat suppreesed, SE spin-echo, GRE gradient-recalled echo, SPIR spectral presaturation inversion recovery, sag obl sagittal oblique, cor obl coronal oblique, cor coronal, sag sagittal, Ax Axial, fs fat-saturated, DE dual-echo, NA not available.

Diagnostic value of MRA

In total, eight studies8,1319 evaluated the performance of MRA to diagnose articular-sided partial-thickness RCTs. The pooled sensitivity, specificity, PLR, NLR, DOR and AUC were 0.81 (95% CI, 0.65–0.90), 0.96 (95% CI, 0.91–0.98), 13.94 (95% CI, 7.25–26.80), 0.23 (95% CI, 0.12–0.44), 68.14 (95% CI, 33.20–139.84) and 0.96 (95% CI, 0.94–0.97), respectively (Figures 1A and 2A, Table 2). TheI 2 statistics for sensitivity and specificity values were 77.19% (95% CI, 61.56%-92.81%) and 88.15% (95% CI, 81.33%-94.97%), respectively, indicating substantial heterogeneity among the included studies. The results of meta-regression analysis revealed that patient sample size, publication year, magnetic field strength, the number of readers, design of the study, blinding, the results of QUADAS-2 score and muscle-tendon did not account for the heterogeneity of sensitivity and specificity and the Deeks' Funnel Plot Asymmetry Test revealed no publication bias (p-value = 0.61) (Figure 3A).

Figure 1.

Figure 1.

The pooled sensitivity, specificity, and summarized receiver operating characteristic curve (SROC) of MRA (A) and MRI (B) to diagnose articular-sided partial-thickness rotator cuff tears with the corresponding 95% confidence region.

Figure 2.

Figure 2.

Forest plots of the pooled diagnostic odds ratio (DOR) of MRA (A) and MRI (B) to diagnose articular-sided partial-thickness rotator cuff tears with the corresponding 95% confidence region.

Table 2.

The accuracy of MRA and MRI for detection of articular-sided partial-thickness rotator cuff tears.

Study Test Sensitivity (95% CI) Specificity (95% CI) PLR (95% CI) NLR (95% CI) DOR (95% CI) AUC (95% CI)
Choo, 2012 3.0-T MRA 0.92 (0.62, 1.00) 0.92 (0.79, 0.98) 11.61 (3.87, 34.86) 0.09 (0.01, 0.59) 128.33 (12.09, 1362.4)
Choo, 2015 3.0-T MRA 0.90 (0.73, 0.98) 0.83 (0.74, 0.90) 5.17 (3.30, 8.10) 0.13 (0.04, 0.37) 41.29 (11.20, 152.20)
Chun, 2010 1.5-T MRA 0.85 (0.73, 0.93) 0.90 (0.84, 0.94) 8.41 (5.14, 13.75) 0.17 (0.09, 0.31) 50.98 (20.29, 128.09)
Lee, 2014 3.0-T MRA 0.56 (0.31, 0.79) 0.98 (0.95, 1.00) 34.82 (10.53, 115.15) 0.45 (0.27, 0.76) 77.08 (17.70, 335.79)
Lee, 2015 3.0-T MRA 0.75 (0.51, 0.91) 0.98 (0.95, 0.99) 32.55 (13.2, 80.27) 0.26 (0.12, 0.55) 127.20 (33.12, 488.54)
Magee, 2009 3.0-T MRA 0.97 (0.84, 1.00) 1.00 (0.97, 1.00) 227.18 (14.28, 3614.60) 0.05 (0.01, 0.22) 4977.0 (1.9E + 2, 1.3E + 5)
Modi, 2013 3.0-T MRA 0.40 (0.12, 0.74) 0.98 (0.92, 1.00) 18.60 (3.88, 89.14) 0.61 (0.37, 1.02) 30.33 (4.59, 200.33)
Oh, 2009 3.0-T MRA 0.67 (0.22, 0.96) 0.90 (0.74, 0.98) 6.67 (1.98, 22.44) 0.37 (0.12, 1.16) 18.00 (2.26, 143.30)
Pooled estimate 0.81 (0.75, 0.87) 0.95 (0.93, 0.96) 13.94 (7.25, 26.80) 0.23 (0.12, 0.44) 68.14 (33.20, 139.84) 0.96 (0.94, 0.97)
Fritz, 2007 1.5-T MRI 0.82 (0.65, 0.93) 0.98 (0.95, 1.00) 41.93 (15.68, 112.13) 0.19 (0.09, 0.38) 226.13 (59.96, 852.77)
Lee, 2015 3.0-T MRI 0.50 (0.12, 0.88) 0.98 (0.92, 1.00) 23 (4.7, 112.47) 0.51 (0.23, 1.14) 45.00 (5.36, 377.56)
Lo, 2016 1.5-T MRI 0.93 (0.77, 0.99) 0.65 (0.51, 0.78) 2.69 (1.83, 3.96) 0.11 (0.03, 0.41) 25.50 (5.44, 119.64)
Magee, 2009 3.0-T MRI 0.69 (0.5, 0.84) 1 (0.97, 1.00) 162.27 (10.11, 2604.7) 0.32 (0.19, 0.53) 507.86 (28.72, 8980.6)
Perez, 2018 1.5-T MRI 0.25 (0.01, 0.81) 0.67 (0.09, 0.99) 0.75 (0.07, 7.73) 1.13 (0.42, 3.00) 0.67 (0.03, 18.06)
Pooled estimate 0.77 (0.68, 0.85) 0.95 (0.92, 0.97) 11.85 (1.71, 82.19) 0.34 (0.17, 0.67) 47.82 (8.29, 275.89) 0.89 (0.86, 0.92)

MRA magnetic resonance angiography, MRI magnetic resonance imaging, T tesla, PLR positive likelihood ratio, NLR negative likelihood ratio, DOR diagnostic odd ratio, AUC area under curve, CI confidence interval.

Figure 3.

Figure 3.

Graphical display of the results of Deek's test for publication bias of MRA (A) and MRI (B).

Diagnostic value of MRI

In total, five studies12,13,17,20,21 assessed the performance of MRI to diagnose articular-sided PTRCTs. The pooled sensitivity, specificity, PLR, NLR, DOR and AUC were 0.78 (95% CI, 0.65–0.87), 0.97 (95% CI, 0.84–0.99), 11.85 (95% CI, 1.71–82.19), 0.34 (95% CI, 0.17–0.67), 47.82 (95% CI, 8.29–275.89) and 0.89 (95% CI, 0.86–0.92), respectively (Figures 1B and 2B, Table 2). TheI 2 statistics for sensitivity and specificity values were 71.57% (95% CI, 45.21%-97.93%) and 95.17% (95% CI, 92.36%-97.97%), respectively, indicating substantial heterogeneity among the included studies. The Deeks' Funnel Plot Asymmetry Test revealed no publication bias (p-value = 0.18) (Figure 3B).

When p values were calculated, MRA showed statistically significantly increased sensitivity compared with conventional MRI for the articular-sided partial-thickness rotator cuff tears (p = 0.038); however, there was not a statistically significant difference shown between MRI and MRA for specificity in detection of the articular-sided partial-thickness rotator cuff tears.

Discussion

This meta-analysis involving eight studies8,1319 for MRA and five studies12,13,17,20,21 for MRI, reveals that MRA has a better diagnostic value than that of MRI for the diagnosis of articular-sided partial-thickness rotator cuff tears because of an improvement of sensitivity (0.81 vs. 0.78).

It has become a popular topic whether contrast agent is used in the application of MRI to diagnose RCTs. A network meta-analysis of 144 diagnostic studies involving 14,059 patients (14,212 shoulders) demonstrated that for the detection of full-thickness tears, partial-thickness tears or any tear, MRA had the highest sensitivity, specificity and superiority index in three common-used imaging modalities (MRA, MRI and US), which revealed that high-field MRA had the highest diagnostic value for detecting any tear, followed by low-field MRA, high-field MRI, high-frequency US, low-field MRI, and low-frequency US. 6 In addition, another meta-analysis of 14 studies involving 1216 patients with labral lesions revealed that MRA had the highest sensitivity and specificity compared with those of MRI and CTA, which indicated that MRA was suggested for use in patients with chronic shoulder symptoms or a pathological abnormality. 22 MRA has a good diagnostic performance mainly depending on the objective evidence of the leakage of contrast agent, accompanied by a good anatomic resolution and subtle defects depicted by contrast agents.23,24 Nonetheless, MRA, as an invasive diagnosis modality, has many disadvantages and limitations, including a longer examination time, increased risk of infection and adverse complications.25,26

MRI, as a non-invasive and reproducible diagnosis image, was recommended for the diagnosis of a patient with suspected rotator cuff injuries. 27 Our lab's research 6 confirmed that the diagnostic value rank (from high to low) of commonly used modalities was 3.0-T MRA, 1.5-T MRA, 3.0-T MRI, 7.5-MHz US, 1.5-T MRI and < 7.5-MHz US in the diagnosis of full-thickness tears; however, in the diagnosis of partial-thickness tears, the diagnostic value rank (from high to low) was 3.0-T MRA, 3.0-T MRI, 1.5-T MRA, 7.5-MHz US, 1.5-T MRI and < 7.5-MHz US. Moreover, other research performed in our lab's confirmed that MRI had similar sensitivity and specificity to MRA in the detection of bursal-sided PTRCTs. 9 In addition, another meta-analysis revealed that MRI is by far the first choice recommendation for imaging modality for the detection of acute labral lesions. 22 Recently, one study in 2020 using hierarchical summary receiver operating characteristic curves (HSROC) demonstrated that MRI was recommended to be a first-choice imaging modality for the detection of rotator cuff tears. 28 Although MRA has a higher sensitivity and specificity, it cannot replace MRI after the comprehensive consideration of accuracy and practicality.

In this study, MRA has a sensitivity of 0.81 and specificity of 0.96 for detecting articular-sided PTRCTs; however, MRI shows a sensitivity of 0.78 and specificity of 0.97 in the detection of the articular-sided PTRCTs. The differences in specificity and sensitivity between MRA and MRI may be a good reason to avoid the potential risk/cost of MRA. This result is in accordance with a prospective study that the sensitivity and specificity of MRA were improved only by 3% to 4% when compared with MRI. 26

Several limitations in this research should merit consideration. At first, the number of patients in the included studies is very small and most of the included studies are retrospective type. Evidence of heterogeneity in the pooled data existed across the included studies. We used the meta-regression analysis to explore many factors that may bring the heterogeneity, including the patient sample size, publication year, magnetic field strength, number of readers, blinding and QUADAS-2 score, design of the study (prospective or retrospective) as well as muscle-tendon were used as covariates. Even though we found the source of heterogeneity, the insufficient data cannot support us to conduct relevant subgroup analyses, which will reduce statistical efficacy power. Furthermore, the field density is very important for the diagnostic value of MRI or MRA in the detection of articular-sided PTRCTs; however, the insufficient data about the field density of MRA/MRI cannot support us to analyze this important influence. We assessed only the diagnostic value of the imaging modalities alone (MRA or MRI) without physical tests. Finally, the safety and cost-effectiveness of MRA/MRI in clinical practice should be assessed systematically.

Conclusion

This meta-analysis reveals that MRA has a better diagnostic value than that of MRI for the diagnosis of articular-sided partial-thickness rotator cuff tears because of an improvement of sensitivity.

Acknowledgements

We thank Research Square to provide a platform that this article has been accepted for publication as a pre-print (DOI: 10.21203/rs.3.rs- 53384/v1).

Author biographies

Lianxin Li is a professor of orthopaedic surgery at Shandong Provincial Hospital affiliated to Shandong First Medical University. His work focuses on the disease of shoulder and elbow.

Jinlei Dong is an associate professor of orthopaedic surgery at Shandong Provincial Hospital affiliated to Shandong First Medical University. His work focuses specifically on the disease of shoulder and elbow.

Qinghu Li is a research fellow of orthopaedic surgery at Shandong Provincial Hospital affiliated to Shandong First Medical University. His work focuses specifically on the treatment of osteomyelitis.

Jinye Dong a research fellow of ultrasound at Weifang People's Hospital. His work focuses on the diagnostic value of ultrasound and medical images for the disease of shoulder and elbow.

Bingzhi Wang is a master student of orthopaedic surgery at Shandong Provincial Hospital affiliated to Shandong First Medical University. His work focuses specifically on the disease of the shoulder and elbow.

Dongsheng Zhou is a professor of orthopaedic surgery at Shandong Provincial Hospital affiliated to Shandong First Medical University. His work focuses specifically on the treatment of pelvic fracture.

Fanxiao Liu is a research fellow of orthopaedic surgery at Shandong Provincial Hospital affiliated to Shandong First Medical University. His work focuses specifically on the diagnosis and treatment of rotator cuff tears. He has worked on many other topics including aseptic loosening of prosthesis and osteolysis.

Footnotes

Funding: This study was supported by the China Scholarship Council (CSC) (Fanxiao Liu, grant number 201808080126), the incubation fund of Shandong Provincial Hospital (Fanxiao Liu, grant number 2020FY019), the Young Scholars Program of Shandong Provincial Hospital, and the Natural Science Foundation of Shandong Province (Fanxiao Liu, grant number ZR202102180575).

Availability of data and materials: All data generated or analyzed during this study are included in this published article.

Authors’ contributions: LFX, DJL and LLX conceived and designed the study. LFX, DJL and LLX performed the search, extraction of data, and methodological assessment. All authors analyzed the data and wrote the paper. All authors read and approved the final manuscript.

Ethics approval and consent to participate: Not applicable.

Consent for publication: All authors read the final manuscript and approved for publication.

Competing interests: The authors declare that they have no competing interests.

References

  • 1.Codman E. The shoulder: rupture of the supraspinatus tendon and other lesionsin or about the subacromial bursal. Malabar (Florida): RE Kreiger, 1934[ISBN:0898747317]. [Google Scholar]
  • 2.Ellman H. Diagnosis and treatment of incomplete rotator cuff tears. Clinical Orthopaedics and Related Research 1990; 254: 64–74. [PubMed] [Google Scholar]
  • 3.Fukuda H. The management of partial-thickness tears of the rotator cuff. The Journal of Bone and Joint Surgery British Volume 2003; 85: 3–11. [DOI] [PubMed] [Google Scholar]
  • 4.Lehman RC, Perry CR. Arthroscopic surgery for partial rotator cuff tears. Arthroscopy: the Journal of Arthroscopic & Related Surgery : Official Publication of the Arthroscopy Association of North America and the International Arthroscopy Association 2003; 19: E81–E84. [DOI] [PubMed] [Google Scholar]
  • 5.Teng A, Liu F, Zhou Det al. et al. Effectiveness of 3-dimensional shoulder ultrasound in the diagnosis of rotator cuff tears: a meta-analysis. Medicine 2018; 97: e12405. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Liu F, Dong J, Shen WJet al. et al. Detecting rotator cuff tears: a network meta-analysis of 144 diagnostic studies. Orthopaedic Journal of Sports Medicine 2020; 8: 2325967119900356. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Breazeale NM, Craig EV. Partial-thickness rotator cuff tears. Pathogenesis and treatment. The Orthopedic Clinics of North America 1997; 28: 145–155. [DOI] [PubMed] [Google Scholar]
  • 8.Chun KA, Kim MS, Kim YJ. Comparisons of the various partial-thickness rotator cuff tears on MR arthrography and arthroscopic correlation. Korean Journal of Radiology 2010; 11: 528–535. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Huang T, Liu J, Ma Yet al. et al. Diagnostic accuracy of MRA and MRI for the bursal-sided partial-thickness rotator cuff tears: a meta-analysis. Journal of Orthopaedic Surgery and Research 2019; 14: 436. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.McInnes MDF, Moher D, Thombs BD, et al. Preferred reporting items for a systematic review and meta-analysis of diagnostic test accuracy studies: the PRISMA-DTA statement. JAMA 2018; 319: 388–396. [DOI] [PubMed] [Google Scholar]
  • 11.Whiting PF, Rutjes AW, Westwood ME, et al. and Group Q. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Annals of Internal Medicine 2011; 155: 529–536. [DOI] [PubMed] [Google Scholar]
  • 12.Fritz LB, Ouellette HA, O’Hanley TAet al. et al. Cystic changes at supraspinatus and infraspinatus tendon insertion sites: association with age and rotator cuff disorders in 238 patients. Radiology 2007; 244: 239–248. [DOI] [PubMed] [Google Scholar]
  • 13.Magee T. 3-T MRI of the shoulder: is MR arthrography necessary? AJR American Journal of Roentgenology 2009; 192: 86–92. [DOI] [PubMed] [Google Scholar]
  • 14.Oh DK, Yoon YC, Kwon JWet al. et al. Comparison of indirect isotropic MR arthrography and conventional MR arthrography of labral lesions and rotator cuff tears: a prospective study. AJR American Journal of Roentgenology 2009; 192: 473–479. [DOI] [PubMed] [Google Scholar]
  • 15.Choo HJ, Lee SJ, Kim OHet al. et al. Comparison of three-dimensional isotropic T1-weighted fast spin-echo MR arthrography with two-dimensional MR arthrography of the shoulder. Radiology 2012; 262: 921–931. [DOI] [PubMed] [Google Scholar]
  • 16.Modi CS, Karthikeyan S, Marks Aet al. et al. Accuracy of abduction-external rotation MRA versus standard MRA in the diagnosis of intra-articular shoulder pathology. Orthopedics 2013; 36: e337–e342. [DOI] [PubMed] [Google Scholar]
  • 17.Lee JH, Yoon YC, Jee Set al. et al. Comparison of three-dimensional isotropic and two-dimensional conventional indirect MR arthrography for the diagnosis of rotator cuff tears. Korean Journal of Radiology 2014; 15: 771–780. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Choo HJ, Lee SJ, Kim JHet al. et al. Delaminated tears of the rotator cuff: prevalence, characteristics, and diagnostic accuracy using indirect MR arthrography. AJR American Journal of Roentgenology 2015; 204: 360–366. [DOI] [PubMed] [Google Scholar]
  • 19.Lee JH, Yoon YC, Jung JYet al. et al. Rotator cuff tears noncontrast MRI compared to MR arthrography. Skeletal Radiology 2015; 44: 1745–1754. [DOI] [PubMed] [Google Scholar]
  • 20.Lo HC, Hung ST, Kuo DPet al. et al. Quantitative diffusion-weighted magnetic resonance imaging for the diagnosis of partial-thickness rotator cuff tears. Journal of Shoulder and Elbow Surgery 2016; 25: 1433–1441. [DOI] [PubMed] [Google Scholar]
  • 21.Perez JR, Massel D, Barrera CMet al. et al. Rotator cuff tears in the pediatric population: comparing findings on arthroscopic evaluation to pre-operative magnetic resonance imaging. Journal of Clinical Orthopaedics and Trauma 2018; 9(Suppl 1): S123–S128. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Liu F, Cheng X, Dong Jet al. et al. Imaging modality for measuring the presence and extent of the labral lesions of the shoulder: a systematic review and meta-analysis. BMC musculoskeletal Disorders 2019; 20: 487. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Jung JY, Jee WH, Chun CWet al. et al. Diagnostic performance of MR arthrography with anterior trans-subscapularis versus posterior injection approach for subscapularis tendon tears at 3.0T. European Radiology 2017; 27: 1303–1311. [DOI] [PubMed] [Google Scholar]
  • 24.Chung CB, Dwek JR, Feng Set al. et al. MR Arthrography of the glenohumeral joint: a tailored approach. AJR American Journal of Roentgenology 2001; 177: 217–219. [DOI] [PubMed] [Google Scholar]
  • 25.Fotiadou AN, Vlychou M, Papadopoulos Pet al. et al. Ultrasonography of symptomatic rotator cuff tears compared with MR imaging and surgery. European Journal of Radiology 2008; 68: 174–179. [DOI] [PubMed] [Google Scholar]
  • 26.Co S, Bhalla S, Rowan Ket al. et al. Comparison of 2- and 3-dimensional shoulder ultrasound to magnetic resonance imaging in a community hospital for the detection of supraspinatus rotator cuff tears with improved worktime room efficiency. Canadian Association of Radiologists Journal = JournalAssociation Canadienne des Radiologistes 2012; 63: 170–176. [DOI] [PubMed] [Google Scholar]
  • 27.Nelson MC, Leather GP, Nirschl RPet al. et al. Evaluation of the painful shoulder. A prospective comparison of magnetic resonance imaging, computerized tomographic arthrography, ultrasonography, and operative findings. The Journal of Bone and Joint Surgery American Volume 1991; 73: 707–716. [PubMed] [Google Scholar]
  • 28.Liu F, Cheng X, Dong Jet al. et al. Comparison of MRI and MRA for the diagnosis of rotator cuff tears: a meta-analysis. Medicine (Baltimore 2020; 99: e19579. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Science Progress are provided here courtesy of SAGE Publications

RESOURCES