Table 2.
Procedural details
Procedural detail | N = 272 |
---|---|
Technical success | 269 (98.9) |
Implanted WATCHMAN size, mm | |
21 | 28 (10.3) |
24 | 48 (17.6) |
27 | 80 (29.4) |
30 | 63 (23.2) |
33 | 50 (18.4) |
Failed implantation | 3 (1.1) |
Total procedural time, min | 89.12 (± 31.86) |
Fluoroscopy time, min | 13.66 (± 8.39) |
Total contrast, mL | 84.66 (± 52.30) |
Length of hospital stay, d | 1.64 (± 3.32) |
Number of devices attempted | 1.30 (± 0.60) |
Baseline TEE performed | 259 (95.2) |
LAA dimension by TEE, mm | 22.35 (± 3.96) |
LAA depth by TEE, mm | 30.37 (± 6.43) |
Baseline CTA performed | 160 (58.8) |
LAA dimension by CCTA, mm | 25.09 (± 4.65) |
LAA depth by CTA, mm | 30.47 (± 6.80) |
Values are n (%), or mean (± standard deviation). WATCHMAN (Boston Scientific, Marlborough, MA).
CCTA, cardiac CTA; CTA, computed tomography angiography; LAA, left atrial appendage; SD, standardized difference; TEE, transesophageal echocardiography.