Table 3. Clinically relevant dimensions obtained from measurements on the micro-CT and OCT volumes of the cadaveric middle ear [46]. ISJ, incudostapedial joint; TM, tympanic membrane; RMSE, root mean squared error.

Measurement ISJ to Umbo ISJ to Promontory Umbo to Promontory ISJ to TM
Modality OCT Micro-CT OCT Micro-CT OCT Micro-CT OCT Micro-CT
Mean ( mm ) 2.813 2.798 2.007 1.999 1.610 1.577 1.869 NA
Standard deviation ( μm ) 36 51 44 64 38 79 27 NA
RMSE ( μm ) 60 52 75 NA